Terumo HX2 Temperature Management System Recalled for Valve Malfunction
Terumo is recalling 103 HX2 Temperature Management Systems because a malfunctioning mixing valve can cause loss of cooling or heating capability during surgery.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used during surgery where loss of temperature control poses a risk of harm, but the source text does not report any injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.
Plain-English summary
Terumo Cardiovascular Systems Corporation is recalling 103 units of the HX2 Temperature Management System. This device provides temperature control for patient blood and cardioplegia solution during cardiovascular surgery. The recall covers units distributed nationwide in the United States and in Japan.
The recall is due to a malfunctioning mixing valve that can result in the loss of cooling and/or heating capability. When this occurs, the system goes to standby mode and displays error messages EOd or EOE, indicating that the valve is unable to move or complete a motion. The mixing valves require replacement at regular intervals.
Healthcare facilities should stop using the affected units and contact Terumo Cardiovascular Systems Corporation for instructions on replacing the mixing valves. The affected units are identified by specific catalog numbers and serial numbers listed in the official recall notice.
The recalled product
- Product
- Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Affected units
- 103
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (10)
- HX2 Temperature Management System
- 120V Catalog # 809810 serial numbers: 00016
- 0001 1116 HX2 Temperature Management System
- 100V Catalog # 812231 serial numbers: 0001
- 1002
- PT0003
- PT0004 HX2 Temperature Management System
- 240V Catalog # 812240 serial numbers: 0001
- 0002 HX2 Temperature Management System
- Trial Kit Catalog # 809810TK serial numbers: 1001 1042
Distribution
Related recalls
Same manufacturer · Terumo Cardiovascular Systems Corporation
See all →- HighCDI OneView System BPM Probe temperature sensor measurement inaccuracy
FDA (Devices) · 2025-07-02
- HighTerumo CDI OneView Monitor Probe May Fail to Display Oxygen Saturation Data
FDA (Devices) · 2025-04-09
- HighCardiovascular Procedure Kits Recalled for Pressure Relief Valve Defect
FDA (Devices) · 2024-04-10
- HighCAPIOX FX25 Oxygenator Recalled for Potential Leaks from Improper Bonding
FDA (Devices) · 2023-04-12
- HighTerumo Recalls CAPIOX NX19 Oxygenator Due to Improper Bonding
FDA (Devices) · 2023-04-12
Same hazard · malfunction
See all →- HighFresenius BlueStar Hemodialysis Machines Recalled for Alarm Defects
FDA (Devices) · 2026-09-02
- ModerateOlympus 4K UHD LCD Monitors Recalled Due to Potential Connector Faults
FDA (Devices) · 2026-07-29
- High
- ModeratePhilips Azurion 3M15 Medical Imaging Systems Recalled Due to Software Issue
FDA (Devices) · 2026-06-10
- SevereElectric Start Pressure Washers Recalled for Carbon Monoxide Hazard
CPSC · 2026-05-14