The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0973-2026·Announced 2025-12-31

Femoral nails, 440 mm right trochanteric, recalled over undersized shaft diameter

Zimmer is recalling 22 Affixus Antegrade Femoral Nails, model 815509440, because an undersized distal shaft diameter could lead to implant fatigue fracture.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: the firm states an undersized shaft could cause implant fatigue fracture requiring surgical intervention, and no injuries are reported.

Plain-English summary

The Affixus Antegrade Femoral Nail, model 815509440, described in the record as Affixus 9 mm, Length 440 mm, Right, Trochanteric Nail, is being recalled by Zimmer, Inc.

Affected implants may have an undersized distal diameter along the length of the shaft, which the firm states could lead to implant fatigue fracture. Zimmer lists the potential health risks if an implant fatigue fracture occurs as adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

The record lists 22 units, carrying UDI-DI (01)00887868143588(17)340820(10)66717557 from lot 66717557. Distribution was nationwide in the United States, in Alabama, Arizona, California, Florida, Indiana, Kentucky, Massachusetts, Missouri, Montana, North Carolina, New Mexico, Rhode Island, Texas and Utah. The FDA has classified the action as Class II.

The enforcement record reports no injuries. Facilities should quarantine affected stock and follow Zimmer's instructions.

The recalled product

Product
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509440 Software Version: N/A Product Description: Affixus 9 mm, Length 440 mm, Right, Trochanteric Nail Component: N/A
Manufacturer
Zimmer, Inc.
Affected units
22

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • Lot Code: Model No 815509440
  • UDI-DI (01)00887868143588(17)340820(10)66717557
  • Lot Number 66717557 Model No 815509440
  • UDI-DI (01)00887868143588(17)350422(10)67101279
  • Lot Number 67101279 Model No 815509440
  • UDI-DI (01)00887868143588(17)350419(10)67101280
  • Lot Number 67101280

Distribution

Distributed nationwide across the United States.

The notice also names 14 states specifically:

Part of a larger recall action

This product is one of 17 recalled by Zimmer, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 17 products in this recall →

Similar recalls