The Recall Desk
Agency classClass IIOur severity:HighFDA (Food)·H-0329-2026·Announced 2025-12-31

Femoral nails, 380 mm left trochanteric, recalled over undersized shaft diameter

Zimmer is recalling 24 Affixus Antegrade Femoral Nails, model 815609380, because an undersized distal shaft diameter could lead to implant fatigue fracture.

The agency’s classification

FDA (Food): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: the firm states an undersized shaft could cause implant fatigue fracture requiring surgical intervention, and no injuries are reported.

Plain-English summary

The Affixus Antegrade Femoral Nail, model 815609380, described in the record as Affixus 9 mm, Length 380 mm, Left, Trochanteric Nail, is being recalled by Zimmer, Inc.

Affected implants may have an undersized distal diameter along the length of the shaft, which the firm states could lead to implant fatigue fracture. Zimmer lists the potential health risks if an implant fatigue fracture occurs as adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

The record lists 24 units, carrying UDI-DI (01)00887868144509(17)340826(10)66717604 from lot 66717604. Distribution was nationwide in the United States, in Alabama, Arizona, California, Florida, Indiana, Kentucky, Massachusetts, Missouri, Montana, North Carolina, New Mexico, Rhode Island, Texas and Utah. The FDA has classified the action as Class II.

The enforcement record reports no injuries. Facilities should quarantine affected stock and follow Zimmer's instructions.

The recalled product

Product
Ritz Peanut Butter Cracker Sandwiches, Net Wt. 27.6 oz.- 20 Count (20 x 1.38-oz. 6-pack carton)
Affected units
70

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Lot Codes: 44000 07584 2

Distribution

Distributed in 8 states:

Similar recalls