Ecolab Recalls Microtek Medical Bur Hole Probe Covers for Endotoxin Testing
Ecolab Inc is recalling specific lots of Microtek Medical Bur Hole Probe Covers because they did not undergo proper endotoxin testing.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a critical safety test (endotoxin) was not performed to proper levels, creating a risk of harm, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Ecolab Inc is recalling 12,546 individual units of Microtek Medical Bur Hole Latex Free Probe Covers with Gel. The affected products are sterile, ethylene oxide (EO) sterilized, and feature IsoSilk. The recall includes 16 specific lot numbers, including 126429V, 126602V, and 137817V.
The recall was initiated because certain lots of these probe covers did not undergo endotoxin (LAL) testing to the proper levels. Endotoxin testing is a standard procedure to ensure that medical devices are free from bacterial contamination that could cause adverse reactions in patients.
The affected products were distributed nationwide in the United States and internationally to countries including Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada. Users should check their inventory for the specific lot numbers listed in the recall notice and stop using any affected units.
The recalled product
- Product
- MICROTEK(TM) MEDICAL BUR HOLE LATEX FREE PROBE COVER W/ GEL, featuring IsoSilk(R), 13cm x 244 cm, 5" x 96", REF PC3688EU, STERILE EO, QTY 20, Microtek Medical, Inc., Columbus, MS 39702
- Manufacturer
- Ecolab Inc
- Affected units
- 12,546
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (16)
- Lot Numbers: 126429V
- 126602V
- 126611V
- 126612V
- 126613V
- 126614V
- 127176V
- 127182V
- 127331V
- 127382V
- 137533V
- 137826V
- 137967V
- 137666V
- 137787V
- and 137817V
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 6 recalled by Ecolab Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 6 products in this recall →Related recalls
Same manufacturer · Ecolab Inc
See all →- HighEcolab Recalls Slush and Warmer Disc Drapes Due to Sterile Barrier Risk
FDA (Devices) · 2021-03-03
- ModerateEcolab Recalls Hand Sanitizer Due to Discoloration and Foreign Substance
FDA (Drugs) · 2020-09-02
- ModerateEcolab Recalls QUIK-CARE Hand Sanitizer Due to Incorrect Alcohol Ingredient
FDA (Drugs) · 2019-06-26
- ModerateEcolab Recalls Express Gel Hand Sanitizer Due to Chemical Contamination
FDA (Drugs) · 2019-04-03
- ModerateEcolab Moisturizing Gel Hand Sanitizer Recalled for Chemical Contamination
FDA (Drugs) · 2019-04-03
Same hazard · endotoxin
See all →- HighMedline Neurological Surgery Kits Recalled for Endotoxin Variability
FDA (Devices) · 2026-09-30
- HighMedline Craniotomy Kits Recalled for Potential Endotoxin Variability
FDA (Devices) · 2026-09-30
- SevereMedline Neurological Sponges Recalled for Potential Endotoxin Variability
FDA (Devices) · 2026-09-30
- HighMedline Neurological Surgery Kits Recalled for Endotoxin Variability
FDA (Devices) · 2026-09-30
- HighMedline Neurological Sponges Recalled for Potential Endotoxin Variability
FDA (Devices) · 2026-09-30
Similar recalls
- LowEcolab Microtek Rolled Probe Cover Recalled for Incorrect Instructions
FDA (Devices) · 2018-07-11
- HighCustom Medical Specialties Recalls Ultrasound Probe Covers for Missing Endotoxin Testing
FDA (Devices) · 2014-02-19
- LowEcolab Microtek Rolled Probe Cover Recalled for Incorrect Expiration Date Labeling
FDA (Devices) · 2016-12-21
- HighMicrotek Medical Recalls Laparoscopic Tubing Sets Due to Unlabeled Latex Packaging
FDA (Devices) · 2015-04-08
- ModerateMicrotek Medical Smoke Evacuator Tube Sets Recalled for Latex Packaging
FDA (Devices) · 2015-04-08
- ModerateEcolab Recalls Microtek Angiography Drapes Due to Packaging Defects
FDA (Devices) · 2016-09-14