Illumina NextSeq 550Dx DNA Analyzer Recalled for Short Circuit Risk
Illumina is recalling 674 NextSeq 550Dx DNA sequence analyzers worldwide. A short circuit may occur after reconnecting circuit components, causing the device to become inoperable.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II medical device with a known functional defect that renders the analyzer inoperable. Although no injuries have been reported, the device's critical role in genetic diagnosis represents a risk-of-harm situation where diagnostic disruption is possible.
Plain-English summary
Illumina, Inc. is recalling 674 NextSeq 550Dx DNA sequence analyzers (Model REF: 20005715, CE, IVD) due to a potential short circuit. The short circuit may occur shortly after disconnecting and reconnecting the circuit board or imaging module if the ribbon cable is not sufficiently reconnected. When this occurs, the analyzer becomes inoperable.
The affected analyzers were distributed in the United States across multiple states and internationally to 27 countries including Australia, Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Vietnam. The recall involves lot numbers NDX550110 through NDX550843.
Users should contact Illumina, Inc. to determine if their analyzer is among the affected units and to arrange inspection and corrective action. Additional information is available from the FDA using recall number Z-0854-2022.
The recalled product
- Product
- Illumina NextSeq 550Dx, REF: 20005715, CE, IVD
- Manufacturer
- Illumina, Inc.
- Affected units
- 674
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- Lot #s NDX550110 - NDX550843
- UDI: (01)00816270020125
Distribution
Related recalls
Same manufacturer · Illumina, Inc.
See all →- HighDNA sequencing reagent kits recalled over flow cell gasket leaks
FDA (Devices) · 2025-12-10
- ModerateFDA Recalls MiSeq Dx Instruments for Software Cybersecurity Vulnerability
FDA (Devices) · 2023-06-28
- HighNextSeq 550Dx sequencing instruments recalled for cybersecurity software vulnerability
FDA (Devices) · 2023-06-28
- HighIllumina NextSeq 550 Genetic Sequencer Cybersecurity Vulnerability Recall
FDA (Devices) · 2022-06-22
- HighIllumina MiSeq Dx Sequencing Instruments Recalled Due to Cybersecurity Vulnerability
FDA (Devices) · 2022-06-22
Same hazard · short circuit
See all →- Severe
- SevereNissan Recalls 2026 Leaf Vehicles Due to High Voltage Battery Fire Risk
NHTSA · 2026-03-25
- SevereVolvo EX30 high voltage battery may short circuit and overheat
NHTSA · 2026-02-20
- SevereFord Escape and Lincoln Corsair plug-in hybrid battery cells may short
NHTSA · 2026-02-17
- HighWestern Star battery to starter stud may short circuit