Stryker Endoscopy Recalls Mini Metzenbaum Scissors Due to Sterilization Labeling Error
Stryker Endoscopy is recalling 180,573 units of Mini Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text explicitly states no illnesses or injuries are reported, and the hazard is a documentation/labeling issue regarding sterilization parameters rather than a direct physical defect causing harm.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of PKG, 5MM INSERT, MINI METZENBAUM SCISSORS, CURVED, 45CM (P/N 0250080574). These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue and surgical auxiliaries. The recall covers units distributed nationwide in the United States and internationally to numerous countries including Canada, the United Kingdom, Germany, and Japan.
The recall was initiated because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. This documentation discrepancy may affect the proper sterilization of the devices.
Healthcare facilities and distributors should stop using the affected instruments and contact Stryker Endoscopy for further instructions. The source text does not report any illnesses or injuries associated with this recall.
The recalled product
- Product
- PKG, 5MM INSERT, MINI METZENBAUM SCISSORS, CURVED, 45CM, P/N 0250080574 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fre
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- 841953 841956
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · mis labeling
See all →- ModerateMedline Trach Tube Kit Recalled for Component Labeling Discrepancy
FDA (Devices) · 2026-09-02
- ModerateAVID Medical Recalls Halyard Total Joint Kits for Latex Labeling Deficiency
FDA (Devices) · 2026-09-02
- ModerateOlympus Recalls PowerSeal Sealer Due to Incorrect Outer Box Labeling
FDA (Devices) · 2026-09-02
- HighSura Tanmen Ramen Noodles Recalled for Undeclared Fish Allergens
FDA (Food) · 2026-09-02
- SevereWhite Vein Advanced Alkaloids Chewable Tablets Recall for Excess Active Ingredient
FDA (Food) · 2026-09-02