IV set 490343 recalled over piggyback backflow and priming occlusion
B. Braun Medical is recalling 1,272 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: medication backflow between IV containers can alter what a patient receives, and no injuries are reported.
Plain-English summary
IV Administration Sets used in gravity IV administration and with B. Braun's Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump are being recalled, including catalog number 490343.
The recall follows the potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers, and the inability to prime the set due to occlusion.
The recall covers 1,272 units carrying primary UDI-DI 04046964593659. Distribution was worldwide, including the United States and internationally to Canada, Germany, Guatemala and Singapore. The FDA has classified the action as Class II.
Backflow can move a secondary medication into the primary bag, and a set that cannot be primed cannot be used. The enforcement record reports no injuries. Facilities should quarantine affected lots and follow B. Braun's instructions.
The recalled product
- Product
- IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,2 CARESITE…
- Manufacturer
- B Braun Medical Inc
- Affected units
- 1,272
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- 1. Catalog Number: 490343
- Primary UDI-DI: 04046964593659
- Unit of Dose UDI-DI: 04046964593642.
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 116 recalled by B Braun Medical Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 116 products in this recall →Related recalls
Same manufacturer · B Braun Medical Inc
See all →- ModerateB Braun SAFSITE Injection Site Valve Recall for Leakage Risk
FDA (Devices) · 2026-10-28
- CriticalB. Braun Recalls Sodium Chloride Injection Due to Particulate Matter
FDA (Drugs) · 2026-09-16
- LowB Braun Recalls Discofix Stopcock Sets Due to Incorrect Expiration Date
FDA (Devices) · 2026-09-09
- SevereB. Braun Recalls Lactated Ringer's Injection Due to Particulate Matter
FDA (Drugs) · 2026-09-02
- ModerateB. Braun kits contain Mastisol adhesive with potentially cracking tubing
FDA (Devices) · 2026-06-10
Same hazard · backflow
See all →- HighIV Administration Sets from Windstone Medical for Infusion Pump Systems
FDA (Devices) · 2026-04-29
- ModerateIV Administration Sets and Gravity Infusion Kits Risk Backflow
FDA (Devices) · 2026-04-29
- HighIV set 490385 recalled over medication backflow between IV containers
FDA (Devices) · 2025-12-03
- HighIV set 490315 recalled over backflow between primary and secondary containers
FDA (Devices) · 2025-12-03
- HighIV set 490519 recalled over piggyback backflow and priming occlusion
FDA (Devices) · 2025-12-03
Similar recalls
- HighIV set 490487 recalled over piggyback backflow and occlusion risk
FDA (Devices) · 2025-12-03
- HighAnesthesia set 490469 recalled over backflow and priming occlusion
FDA (Devices) · 2025-12-03
- ModerateIV Administration Sets Potential Backflow and Occlusion Risk
FDA (Devices) · 2026-04-29
- HighIV Infusion Pump Set Recalled for Backflow Risk During Medication Administration
FDA (Devices) · 2024-07-24
- HighIV Administration Sets from Windstone Medical for Infusion Pump Systems
FDA (Devices) · 2026-04-29
- ModerateIV Administration Sets and Gravity Infusion Kits Risk Backflow
FDA (Devices) · 2026-04-29