The Recall Desk

Hazard

Priming Failure recalls

120 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
120
Critical
0
Severe
0
Most recent
2025-12-03

Of the 120 priming failure recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all priming failure recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

1–25 of 120

  • HighFDA (Devices)·Z-0645-2026·2025-12-03

    IV set 490067 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 407,800 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 15DR 5CARESITE 1 ULR
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0640-2026·2025-12-03

    Universal infusion sets recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 49,330,370 Infusomat administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0664-2026·2025-12-03

    IV set 490315 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 11,832 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15 DR/ML,2 CA
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0672-2026·2025-12-03

    IV set 490519 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 1,080 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP, 132 I
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0633-2026·2025-12-03

    IV set 490283 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 65,904 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXTENSION SET W/ B/C VALVE 53 in.;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0610-2026·2025-12-03

    Extension set 490358 recalled over backflow into the primary container

    B. Braun Medical is recalling 1,200 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15D, 144 IN.; Cat
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0703-2026·2025-12-03

    IV set 490301 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 31,392 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD.SET W/15 DR 128 IN, E
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0671-2026·2025-12-03

    Vented pump sets recalled over medication backflow and occlusion risk

    B. Braun Medical is recalling 12,168 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0696-2026·2025-12-03

    IV set 490437 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 4,536 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0681-2026·2025-12-03

    IV set 490426 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 63,192 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/10 DROPS, 13
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0687-2026·2025-12-03

    IV set 490407 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 10,896 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/CARESITE EXT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0677-2026·2025-12-03

    IV set 490612 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 5,952 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 15DR 2SAFELINE 1CARESITE 1S
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0658-2026·2025-12-03

    IV set 470005 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 29,900 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET 15DR 3 CARESITE, 124
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0650-2026·2025-12-03

    IV set 490317 recalled over backflow and occlusion risk

    B. Braun Medical is recalling 6,650 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD SET CARESITE 127 IN.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0631-2026·2025-12-03

    Extension set 490629 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 13,128 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXTENSION SET, 25 IN.; Catalog Numb
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0667-2026·2025-12-03

    Infusion pump sets recalled over medication backflow and occlusion

    B. Braun Medical is recalling 37,656 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS, 16
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0634-2026·2025-12-03

    Extension sets recalled over piggyback backflow into primary containers

    B. Braun Medical is recalling 23,650 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXTENSION SET W/ ULTRAPORT, 48 IN.;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0612-2026·2025-12-03

    Anesthesia set 490444 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 88,200 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15D 2 CARESITE LA
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0625-2026·2025-12-03

    Caresite set 354202 recalled over backflow and inability to prime

    B. Braun Medical is recalling 2,871,707 units caresite sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set with 2 CARESITE¿
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0622-2026·2025-12-03

    Burette set 490159 recalled over backflow into the primary container

    B. Braun Medical is recalling 7,240 burette sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. BURETTE SET W/ 2 CARESITE, 134 IN.; C
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0717-2026·2025-12-03

    IV set 490508 recalled over backflow into the primary container

    B. Braun Medical is recalling IV administration set catalog number 490508 over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15DROPS/ML;
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0655-2026·2025-12-03

    IV set 457504 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 80,650 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 2 ULT
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0649-2026·2025-12-03

    IV set 490208 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 40,008 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DR / mL W
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0691-2026·2025-12-03

    IV set 490408 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 24,072 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2 CARESITES,
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-0642-2026·2025-12-03

    IV set 490385 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 168,816 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. IV ADMIN SET, 10 DROPS/ML,
    Category
    Medical Device
    Distribution
    0 states