Agency classClass IIOur severity:HighFDA (Devices)·Z-0646-2026·Announced 2025-12-03

IV set 457506 recalled over piggyback backflow and occlusion risk

B. Braun Medical is recalling 1,019 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: medication backflow between IV containers can alter what a patient receives, and no injuries are reported.

Plain-English summary

IV Administration Sets used in gravity IV administration and with B. Braun's Infusomat Space Large Volume Pump, Outlook Pump and Vista Basic Pump are being recalled, including catalog number 457506.

The recall follows the potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers, and the inability to prime the set due to occlusion.

The recall covers 1,019 units carrying primary UDI-DI 04046964716126. Distribution was worldwide, including the United States and internationally to Canada, Germany, Guatemala and Singapore. The FDA has classified the action as Class II.

Backflow can move a secondary medication into the primary bag, and a set that cannot be primed cannot be used. The enforcement record reports no injuries. Facilities should quarantine affected lots and follow B. Braun's instructions.

The recalled product

Product
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 1…
Affected units
1,019

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • 1. Catalog Number: 457506
  • Primary UDI-DI: 04046964716126
  • Unit of Dose UDI-DI: 04046964716119.

Distribution

Distribution scope not specified by the agency.

Part of a larger recall action

This product is one of 116 recalled by B Braun Medical Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 116 products in this recall →