The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0599-2013·Announced 2012-12-26

Medrad Continuum MR Infusion System Tubing Sets Recalled for Flow Accuracy Issues

Bayer HealthCare is recalling Medrad Continuum MR Infusion System tubing sets because a variation may cause flow rate accuracy issues when used with Continuum pumps.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a performance defect (flow rate accuracy) that poses a risk of harm to patients, but the source text does not report any specific injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Bayer HealthCare is recalling all MEDRAD(R) Continuum MF Infusion System tubing sets, including macrobore and microbore types. The affected catalog numbers are MIK200A, MIL 200B, 3015153, 3015155, 3015156, 3015157, 3015158, 3015159, and 3015161.

The firm is initiating this recall because the tubing exhibits a variation that, when used in conjunction with any Continuum pump, may result in flow rate accuracy performance issues. The product is a peristaltic pump system designed for patients who require medications and other fluids during an MR procedure.

The recall covers 1,229 pumps total with worldwide distribution, including the United States and numerous international locations. Specific serial and lot numbers can be found on the firm's website at www.medrad.com.

The recalled product

Product
Medrad(R) Continuum MR Infusion System Primary VSA Stopcock Set Peristaltic Pump System designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR…
Manufacturer
Medrad Inc
Affected units
1,229

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 11 recalled by Medrad Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 11 products in this recall →

Similar recalls