The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0574-2016·Announced 2016-01-13

Corflex Post-Operative Knee Brace Recalled for Potential Locking Failure

Corflex is recalling 1,406 post-operative knee braces because they may fail to maintain a locked position, potentially causing loss of balance or undesired knee movement.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (potential loss of balance or undesired joint movement) where no specific injuries or illnesses have been reported in the source text, fitting the 'High' severity criteria.

Plain-English summary

Corflex is recalling 1,406 units of the Contender Post-Op Knee Brace Lite (Model 75-7500-000). The device is indicated for range of motion control at the knee joint following various surgeries, including ACL, PCL, MCL, and LCL repairs, meniscal repairs, and total knee replacements.

The recall is due to a potential defect where the knee brace could fail to maintain the locked position. If this occurs, a loss of balance or undesired flexion/extension may happen at the affected knee joint. The affected units were distributed nationwide in the US and internationally to Canada, Taiwan, Singapore, and England.

Consumers should identify affected units by checking the Revision Code. All braces with a Revision Code beginning with AN, AP, AQ, AR, AS, AT, or AU should be returned to Corflex immediately for review. Additionally, braces with a Revision Code beginning with AM that were manufactured between September 1, 2015, and November 6, 2015, are also included in the recall.

The recalled product

Product
Corflex Contender Post-Op Knee Brace Lite Model: 75-7500-000 Post-Operative Knee Brace is indicated for range of motion control at the knee joint following ACL, PCL, MCL, and LCL surgeries, meniscal repairs, patella realignment, regenerative chondroplasty, stable femoral…
Manufacturer
Corflex
Affected units
1,406

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (9)

  • (1) all braces that begin with a Revision Code of AN
  • AP
  • AQ
  • AR
  • AS
  • AT
  • AU are to be returned to Corflex immediately for review
  • 2015 and November 6
  • 2015

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 7 recalled by Corflex under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 7 products in this recall →

Similar recalls