The Recall Desk
HighFDA (Devices)·Z-0445-2018·Announced 2018-02-07

Zimmer Biomet Recalls Affixus Hip Fracture Nails Due to Assembly Defect

Zimmer Biomet is recalling 638 Affixus Hip Fracture Nails because a manufacturing rework process caused set screws and sleeves to shift, potentially preventing proper assembly.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device where a mechanical defect could prevent proper assembly and function, representing a risk of harm even though no specific injuries or illnesses have been reported in the source text.

Plain-English summary

Zimmer Biomet, Inc. is recalling 638 units of the Affixus Hip Fracture Nail, LH 130 DEG 9MM X 400MM. The affected devices were distributed nationwide in the United States, as well as in Portugal and the Netherlands.

The recall was initiated after it was detected during a rework process for a previous field action that the set screw and sleeve assembly, which are designed to have an interference fit, were no longer in their installed locations. This defect could prevent the product from being properly assembled with the mating screw and targeting jig.

Healthcare providers and facilities that have received these specific devices should stop using them and contact Zimmer Biomet for instructions on how to return or replace the affected units.

The recalled product

Product
Affixus Hip Fracture Nail, LH 130 DEG 9MM X 400MM
Affected units
638

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • Item No. 814609400
  • 701940 (06/03/2027) UDI (01)00887868039720 (17)270603 (10)701940
  • 858970 (06/14/2027) UDI (01)00887868039720 (17)270614 (10)858970

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 65 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.