The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0429-2019·Announced 2018-11-21

AROA Biosurgery Recalls OviTex 2S Reinforced BioScaffold Mesh Devices

AROA Biosurgery is recalling OviTex 2S Reinforced BioScaffold 10x12cm devices because the suture material degraded, causing them to fail burst specifications.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device that failed to meet pre-defined burst specifications, representing a risk of harm where injury has not yet been reported, which aligns with the High severity criteria.

Plain-English summary

AROA Biosurgery, LTD. is recalling 881 units of OviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012G. The recall was initiated after an ongoing product stability study observed degradation in the PGA suture material used to manufacture these resorbable mesh devices.

Further investigation revealed that devices older than 18 months showed evidence of no longer meeting the pre-defined ball burst specification relative to the number of tissue layers. The affected products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.

Healthcare providers and patients should stop using the recalled devices and contact AROA Biosurgery, LTD. for further instructions on how to return or dispose of the product.

The recalled product

Product
OviTex 2S Reinforced BioScaffold 10x12cm, Part Number F10258-1012G
Affected units
881

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • ERT-6H16

Distribution

Distributed in 8 states:

Part of a larger recall action

This product is one of 21 recalled by AROA Biosurgery, LTD. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 21 products in this recall →

Similar recalls