Agency classClass IIOur severity:HighFDA (Devices)·Z-0419-2019·Announced 2018-11-21

AROA Biosurgery Recalls OviTex Reinforced BioScaffold Mesh Devices

AROA Biosurgery is recalling OviTex Reinforced BioScaffold 16x20cm devices because the suture material degraded, causing them to fail burst specifications after 18 months.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device that failed to meet pre-defined burst specifications, posing a risk of harm where injury has not yet been reported, which aligns with the High severity criteria.

Plain-English summary

AROA Biosurgery, LTD. is recalling 881 units of OviTex Reinforced BioScaffold 16x20cm, Part Number F10254-1620G. The recall was initiated after an ongoing product stability study observed degradation in the PGA suture material used to manufacture these resorbable mesh devices.

Further investigation indicated that devices older than 18 months showed evidence of no longer meeting the pre-defined ball burst specification relative to the number of tissue layers. The affected products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.

Healthcare providers and patients should stop using the recalled devices and contact AROA Biosurgery for further instructions. The specific lot codes associated with this recall are ERT-6G08 and ERT-6J03.

The recalled product

Product
OviTex Reinforced BioScaffold 16x20cm, Part Number F10254-1620G
Affected units
881

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • ERT-6G08 ERT-6J03

Distribution

Distributed in 8 states: