Breas Medical Ventilator Mouthpiece Recall: Unintended Slits May Reduce Ventilation
Breas Medical is recalling 215 mouthpiece units due to manufacturing defects with unintended slits that may result in insufficient ventilation in mechanical ventilation patients.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a risk-of-harm medical device where serious injury (inadequate ventilation in a ventilator patient) has not yet been reported. No hospitalizations or deaths are documented in the source, and the hazard is stated as theoretical ('may result'). Class II device recall with potential safety impact but no actual harm reported fits the 'High' severity category.
Plain-English summary
Breas Medical, Inc. is recalling the NIV Angled Mouthpiece with Notch 22M/15F, a respiratory accessory used with the Vivo 45LS ventilator. The device is intended for partial insertion into a patient's mouth to facilitate respiratory access. The recall affects 215 units distributed nationwide and in Canada, with lot numbers 32304444, 32305269, and 32306017.
The mouthpiece units were produced with unintended open slits on the side of the mouthpiece body. This manufacturing defect may result in insufficient ventilation in patients using the device in spontaneous breathing modes (MPV-VCV and MPV-PCV ventilation).
Patients and healthcare providers should cease use of affected units and contact Breas Medical or their equipment supplier for replacement or corrective action.
The recalled product
- Product
- with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
- Manufacturer
- Breas Medical, Inc.
- Affected units
- 215
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- UDI-DI: 05030267120505 NIV Angled Mouthpiece with Notch 22M/15F
- 07321820060931 MPV Kit 006093
- 07321820066537 MPV Kit 006653 MPV mouthpiece Lot Numbers: 32304444
- 32305269 and 32306017
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Breas Medical, Inc.
See all →- SevereVivo 45 LS Ventilator Recall: Potential Formaldehyde Emission Risk
FDA (Devices) · 2024-09-11
- HighVivo 45LS Ventilator processor failure alarm not generated on shutdown
FDA (Devices) · 2022-07-20
- HighVentilator trolley mounting brackets can detach and cause device to fall
FDA (Devices) · 2022-04-20
- HighBreas Medical Recalls Vivo Ventilator Protective Covers Due to Obstruction Risk
FDA (Devices) · 2020-03-04
Same hazard · manufacturing defect
See all →- HighBeak & Skiff Alcohol Antiseptic Sanitizer recalled for acetaldehyde contamination
FDA (Drugs) · 2026-10-14
- HighFDA Recalls Opportunity Pro Nitrile Gloves Due to Manufacturing Defects
FDA (Devices) · 2026-10-07
- HighWright Medical Recalls Charlotte Claw Screws Due to Missing Threads
FDA (Devices) · 2026-10-07
- ModerateCustom Interventional Pain Management Recalls Mastisol Skin Adhesive Trays
FDA (Devices) · 2026-10-07
- HighNOxBOXi Nitric Oxide Delivery System Recalled for Oxygen Sensor Defect
FDA (Devices) · 2026-10-07
Similar recalls
- SevereVivo 45 LS Ventilator Recall: Potential Formaldehyde Emission Risk
FDA (Devices) · 2024-09-11
- HighBreas Medical Recalls Vivo Ventilator Protective Covers Due to Obstruction Risk
FDA (Devices) · 2020-03-04
- HighBreas Vivo 50 Home Care Ventilator Recalled for Keypad Malfunction
FDA (Devices) · 2015-07-22
- HighVivo 45LS Ventilator processor failure alarm not generated on shutdown
FDA (Devices) · 2022-07-20
- HighVentilator trolley mounting brackets can detach and cause device to fall
FDA (Devices) · 2022-04-20
- HighMedtronic Arrive Braided Transseptal Sheath Recalled for Inner Liner Fragment
FDA (Devices) · 2016-12-28