The Recall Desk

Manufacturer

Breas Medical, Inc.

5 recalls in our database name Breas Medical, Inc. as the manufacturing or recalling firm.

What the numbers show

Total
5
Critical
0
Severe
1
Most recent
2024-09-11

Of the 5 recalls from Breas Medical, Inc. we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1 (20%) are Severe.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

1–5 of 5

  • SevereFDA (Devices)·Z-2690-2024·2024-09-11

    Vivo 45 LS Ventilator Recall: Potential Formaldehyde Emission Risk

    Breas Medical is recalling the Vivo 45 LS ventilator due to potential elevated formaldehyde emissions into the patient breathing airpath under specific conditions.

    Product
    Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog Number: 230000 (230016, 230216)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0405-2024·2023-12-06

    Breas Medical Ventilator Mouthpiece Recall: Unintended Slits May Reduce Ventilation

    Breas Medical is recalling 215 mouthpiece units due to manufacturing defects with unintended slits that may result in insufficient ventilation in mechanical ventilation patients.

    Product
    with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1382-2022·2022-07-20

    Vivo 45LS Ventilator processor failure alarm not generated on shutdown

    Breas Medical recalls 1,113 Vivo 45LS Ventilators due to a processor that can fail without triggering an alarm, potentially preventing detection of ventilator malfunction if monitoring instructions are not followed.

    Product
    Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0943-2022·2022-04-20

    Ventilator trolley mounting brackets can detach and cause device to fall

    The mounting brackets on Breas Ventilator Trolleys can become detached, allowing ventilators to fall. This poses an injury risk to people near the equipment.

    Product
    Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1352-2020·2020-03-04

    Breas Medical Recalls Vivo Ventilator Protective Covers Due to Obstruction Risk

    Breas Medical is recalling 629 Vivo ventilator protective covers because the design may obstruct the exhalation valve control pressure tube.

    Product
    Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65…
    Category
    Medical Device
    Distribution
    Distributed nationwide