Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns
Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where injury has not yet been reported, and the hazard is theoretical ('cannot rule out'), fitting the High severity criteria.
Plain-English summary
Zimmer Biomet, Inc. is recalling 412 units of the ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm (Number: 00-8065-654-28). The recall covers all lots of this specific product, which was distributed nationwide in the United States.
The recall was initiated because the hydrogen peroxide sterilant used in the acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. As a result, the company cannot rule out the potential for adverse tissue reactions in patients.
This recall affects medical professionals and patients who have received or may receive this specific hip prosthesis component. Consumers and healthcare providers should contact Zimmer Biomet for further instructions on how to proceed with the recall and any necessary medical follow-up.
The recalled product
- Product
- ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-28
- Manufacturer
- Zimmer Biomet, Inc.
- Affected units
- 412
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lots
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 75 recalled by Zimmer Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 75 products in this recall →Related recalls
Same manufacturer · Zimmer Biomet, Inc.
See all →- HighZimmer Biomet Recalls Spinal Rod Cutters Due to Fracture Risk
FDA (Devices) · 2020-12-30
- ModerateZimmer Biomet Recalls Polaris Spinal System Screws for Labeling Update
FDA (Devices) · 2020-12-23
- HighZimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns
FDA (Devices) · 2020-11-11
Same hazard · adverse tissue reaction
See all →- HighCodman Cranial Access Kits Recall: Incorrect Expiration Dates on Labels
FDA (Devices) · 2022-03-30
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns
FDA (Devices) · 2020-11-11
- HighZimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns
FDA (Devices) · 2020-11-11
Similar recalls
- ModerateZimmer Biomet Recalls Bipolar Shell Hip Prostheses Due to Packaging Defect
FDA (Devices) · 2016-03-02
- SevereZimmer Biomet Recalls Modular Hip Head Components Due to Sizing Errors
FDA (Devices) · 2018-05-09
- ModerateZimmer Biomet Ceramic Hip Prosthesis Head Lacks Stem Compatibility Information
FDA (Devices) · 2025-04-30
- ModerateZimmer Biomet Recalls Hip Implants Due to Packaging Residue
FDA (Devices) · 2018-03-21
- SevereZimmer Recalls M/L Taper Hip Stems Due to Cytotoxicity Levels
FDA (Devices) · 2015-06-17
- ModerateZimmer Biomet Ceramic Femoral Head Hip Prosthesis Labeling Recall
FDA (Devices) · 2025-04-30