The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-0364-2027·Announced 2026-11-18

Siemens Atellica CH HDL Cholesterol Reagent Recall for Carryover Issues

Siemens Healthcare Diagnostics is recalling Atellica CH HDL Cholesterol reagents because carryover may cause inaccurate Hemoglobin A1c and Direct Bilirubin results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as Class II and involves reagent carryover that may cause inaccurate Hemoglobin A1c and Direct Bilirubin results. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Siemens Healthcare Diagnostics, Inc. is recalling the Atellica CH HDL Cholesterol (HDLC) reagent, identified by Siemens Material Number (SMN) 11537213. The recall involves 72,207 units distributed worldwide, including the United States and numerous international markets.

The recall is issued because reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover, with biases of 9 to 29% in NGSP units. Additionally, carryover may cause falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover, with biases of 0% to -25%. These inaccuracies may affect patient samples, quality control, and/or calibrator results.

The source text does not specify instructions for consumers or healthcare providers regarding the handling of the recalled product or the management of affected test results.

The recalled product

Product
Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213;

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • Siemens Material Number (SMN): 11537213
  • UDI-DI: 00630414610832
  • All lots

Distribution

Distributed nationwide across the United States.