Siemens Atellica CH Total Protein Reagent Recalled for Carryover Errors

Siemens Healthcare Diagnostics is recalling Atellica CH Total Protein reagents because probe carryover may cause inaccurate Hemoglobin A1c and Direct Bilirubin results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as Class II and involves reagent carryover that may cause inaccurate Hemoglobin A1c and Direct Bilirubin test results. The source does not report any illnesses, injuries, or deaths.

Plain-English summary

Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH Total Protein_2 (TP_2) reagents. The affected products are identified by Siemens Material Numbers 11645755 (3700 tests) and 11537228 (7400 tests). The recall covers 25,210 units distributed worldwide, including 8,087 units in the United States and 17,123 units in other countries.

The recall is due to reagent probe carryover on the Atellica CH and Atellica CI Analyzers. This carryover may cause falsely elevated Hemoglobin A1c results due to HDLC carryover, with biases of 9 to 29% in NGSP units. It may also cause falsely depressed Direct Bilirubin 2 results due to TP_2 carryover, with biases of 0% to -25%. These errors may affect patient samples, quality control, and/or calibrator results.

Healthcare facilities and laboratories using these reagents should stop using the affected products and contact Siemens Healthcare Diagnostics for further instructions. The source text does not report any illnesses, injuries, or deaths associated with this recall.

The recalled product

Product
Atellica CH Total Protein_2 (TP_2); Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (5)

  • 1. Siemens Material Number (SMN): 11645755 (3700 tests)
  • UDI-DI: 00630414654010
  • All lots
  • 2. Siemens Material Number (SMN): 11537228 (7400 tests)
  • UDI-DI: 00630414611082

Distribution

Distributed nationwide across the United States.