Siemens Atellica CH Total Protein Reagent Recalled for Carryover Errors
Siemens Healthcare Diagnostics is recalling Atellica CH Total Protein reagents because probe carryover may cause inaccurate Hemoglobin A1c and Direct Bilirubin results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as Class II and involves reagent carryover that may cause inaccurate Hemoglobin A1c and Direct Bilirubin test results. The source does not report any illnesses, injuries, or deaths.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH Total Protein_2 (TP_2) reagents. The affected products are identified by Siemens Material Numbers 11645755 (3700 tests) and 11537228 (7400 tests). The recall covers 25,210 units distributed worldwide, including 8,087 units in the United States and 17,123 units in other countries.
The recall is due to reagent probe carryover on the Atellica CH and Atellica CI Analyzers. This carryover may cause falsely elevated Hemoglobin A1c results due to HDLC carryover, with biases of 9 to 29% in NGSP units. It may also cause falsely depressed Direct Bilirubin 2 results due to TP_2 carryover, with biases of 0% to -25%. These errors may affect patient samples, quality control, and/or calibrator results.
Healthcare facilities and laboratories using these reagents should stop using the affected products and contact Siemens Healthcare Diagnostics for further instructions. The source text does not report any illnesses, injuries, or deaths associated with this recall.
The recalled product
- Product
- Atellica CH Total Protein_2 (TP_2); Siemens Material Number (SMN): 1. 11645755 (3700 tests), 2. 11537228 (7400 tests);
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- 1. Siemens Material Number (SMN): 11645755 (3700 tests)
- UDI-DI: 00630414654010
- All lots
- 2. Siemens Material Number (SMN): 11537228 (7400 tests)
- UDI-DI: 00630414611082
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- ModerateSiemens Atellica CH HDL Cholesterol Reagent Recall for Carryover Issues
FDA (Devices) · 2026-11-18
- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · inaccurate results
See all →- ModerateRadiometer ABL90 FLEX Analyzers Recalled for Inaccurate Results
FDA (Devices) · 2026-10-21
- SevereInstrumentation Laboratory Recalls HemoslL AcuStar HIT-lgG In Vitro Diagnostic
FDA (Devices) · 2026-10-21
- ModerateBeckman Coulter DxC 500 AU Clinical Chemistry Analyzers Recalled for Software Error
FDA (Devices) · 2026-10-14
- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
Similar recalls
- ModerateSiemens Atellica CH HDL Cholesterol Reagent Recall for Carryover Issues
FDA (Devices) · 2026-11-18
- ModerateSiemens Atellica CH Total Protein II Recalled for Reagent Carryover
FDA (Devices) · 2023-09-06
- HighLaboratory diagnostic test may produce inaccurate results from reagent carryover
FDA (Devices) · 2023-09-06
- HighSiemens Atellica CH Calcium_2 Reagent Carryover Recall
FDA (Devices) · 2022-04-20
- HighSiemens Atellica CH Protein Reagent Recalled for Carryover Affecting Test Results
FDA (Devices) · 2021-11-03
- HighSiemens Atellica CH diagnostic reagent carryover may cause false laboratory test results
FDA (Devices) · 2023-09-06