Pega Medical Recalls SCFE Driver Instruments Due to Threaded Shaft Breakage
Pega Medical Inc. is recalling 4 SCFE Driver instruments distributed in Florida and Tennessee because the threaded shaft broke during use.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device failure (shaft breakage) during a surgical procedure, which presents a risk of harm to patients, but the source text does not report any actual injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Pega Medical Inc. is recalling 4 units of the SCFE Driver (SCF-MLD273 & SCF-MLD265) instruments. These instruments are used for the insertion of the Free Gliding SCFE Screw System, a temporary implant for stabilizing pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in pediatric patients. The recalled instruments are part of the Free Gliding SCFE Screw System instrument tray and are marked with specific catalogue and lot numbers.
The recall was initiated because the threaded shaft of the SCFE Driver broke close to the knob when manual force was applied during the fixation of the screw. The affected units were distributed in the states of Florida and Tennessee. The specific part numbers and lot numbers involved are SCF-MLD265 (Lot 130703-01) and SCF-MLD273 (Lot 130703-02).
Healthcare providers and facilities that have these specific instruments should stop using them immediately and contact Pega Medical Inc. for instructions on return or replacement. The source text does not report any patient injuries or illnesses associated with this failure.
The recalled product
- Product
- SCFE Driver (SCF-MLD273 & SCF-MLD265) instrument used for the insertion of the Free Gliding SCFE Screw System. This instrument is included in the Free Gliding SCFE Screw System instrument tray. This instrument is marked with the catalogue number and lot number. The…
- Manufacturer
- Pega Medical Inc.
- Hazard
- Affected units
- 4
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Part Number: SCF-MLD265
- LOT#: 130703-01 Part Number: SCF-MLD273
- LOT#: 130703-02
Distribution
Related recalls
Same manufacturer · Pega Medical Inc.
See all →- LowPega Medical Recalls SLIM System Handle Due to Lot Number Mismatch
FDA (Devices) · 2020-05-27
- ModeratePega Medical Recalls SCFE Knobs for Driver Due to Screw Loosening Risk
FDA (Devices) · 2019-11-13
- HighPega Medical Recalls SLIM Handle Components Due to Jamming Risk
FDA (Devices) · 2019-08-21
- HighPega Medical Recalls SCFE Drivers Due to Handle Detachment Risk
FDA (Devices) · 2015-07-08
- HighPega Medical Recalls Female Driver Instruments for Orthopedic Procedures
FDA (Devices) · 2015-05-06
Same hazard · mechanical failure
See all →- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- HighPega Medical Recalls SCFE Drivers Due to Handle Detachment Risk
FDA (Devices) · 2015-07-08
- ModeratePega Medical Recalls SCFE Knobs for Driver Due to Screw Loosening Risk
FDA (Devices) · 2019-11-13
- HighPega Medical Recalls Female Driver Instruments for Orthopedic Procedures
FDA (Devices) · 2015-05-06
- HighSpineFrontier Recalls Rev E PedFuse Return Screw Inserters Due to Loose Set Screw
FDA (Devices) · 2013-08-14
- LowPega Medical Recalls SLIM System Handle Due to Lot Number Mismatch
FDA (Devices) · 2020-05-27
- HighPega Medical Recalls SLIM Handle Components Due to Jamming Risk
FDA (Devices) · 2019-08-21