Ortho-Clinical Diagnostics Recalls VITROS Chemistry Cl- Slides Due to Dispense Issues
Ortho-Clinical Diagnostics is recalling VITROS chemistry Cl- Slides because the anti-backup platen may fail to advance, causing intermittent slide dispense issues.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a mechanical malfunction (dispense issues) rather than a direct health risk like contamination or injury, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Ortho-Clinical Diagnostics is recalling VITROS chemistry Products Cl- Slides, Catalog No. 8445207. These MicroSlides contain reagents in a dry, multi-layered form and are used for chemistry testing. The recall involves 1,365,545 total units.
The recall was initiated because users may experience intermittent slide dispense issues with some cartridges. This is attributed to the anti-backup platen failing to advance inside the cartridge. The product was distributed domestically nationwide and in foreign markets including Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, the Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, and the United Kingdom.
Users of these products should check their inventory for the affected catalog number and contact Ortho-Clinical Diagnostics for further instructions on how to proceed with the recall.
The recalled product
- Product
- VITROS chemistry Products Cl- Slides, Catalog No. 8445207 (which does NOT support Urine CL-). MicroSlides which contain reagents in a dry, multi-layered form.
- Manufacturer
- Ortho-Clinical Diagnostics
- Hazard
- Affected units
- 1,365,545
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Ortho-Clinical Diagnostics under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Ortho-Clinical Diagnostics
See all →- HighClinical Laboratory Quality Control Device Recalled for Diluent Assembly Error
FDA (Devices) · 2024-04-03
- HighLaboratory performance verifier recalled for mismatched diluent assembly
FDA (Devices) · 2024-04-03
- HighVITROS Troponin I ES Immunodiagnostic Reagent Pack Recalled for Incorrect Wells
FDA (Devices) · 2023-08-16
- HighFDA Recalls Troponin Cardiac Test Calibrators for Potential False Negatives
FDA (Devices) · 2023-08-16
- HighOrtho Recalls VITROS Chemistry Calibrator Kit 20 for Analytical Bias
FDA (Devices) · 2023-07-19
Same hazard · mechanical failure
See all →- SevereCanon Medical Alphenix INFX-8000C/B5 Recalled for Tabletop Shift Risk
FDA (Devices) · 2026-10-14
- ModerateLeMaitre TufTex Over-the-Wire Embolectomy Catheter Recalled for Balloon Failure
FDA (Devices) · 2026-10-14
- HighMirion Captus 4000e Thyroid Uptake System Recalled for Spring Arm Failure
FDA (Devices) · 2026-09-23
- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
Similar recalls
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Ca Slides Due to Dispense Issues
FDA (Devices) · 2013-11-27
- LowOrtho-Clinical VITROS Chloride Slides Recalled for Interference Issues
FDA (Devices) · 2018-09-19
- LowOrtho-Clinical Diagnostics Recalls VITROS Chemistry Products Cl- Slides
FDA (Devices) · 2018-09-19
- ModerateOrtho-Clinical Diagnostics Recalls VITROS Chemistry Slide Dispense Cartridges
FDA (Devices) · 2013-11-27
- HighOrtho-Clinical Diagnostics Recalls VITROS Chemistry CK Slides for Dispense Issues
FDA (Devices) · 2013-11-27
- ModerateOrtho-Clinical Recalls VITROS Chemistry Slide Dispense Cartridges
FDA (Devices) · 2013-11-27