The Recall Desk
HighFDA (Devices)·Z-0214-2026·Announced 2025-10-29

Central monitoring stations recalled over reboot loop at specific network sizes

GE HealthCare is recalling 2,306 Carescape Central Stations running software V3.0.5 because the system can enter a continuous reboot cycle, losing central monitoring of connected patients.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: a loss of central patient monitoring is a stated clinical risk, and no injuries are reported.

Plain-English summary

The GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, part number 5867474-03, is being recalled.

GE HealthCare has identified an issue that can cause the system to enter a continuous reboot cycle. If that occurs, a loss of central monitoring of connected patients could result. The firm states the issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

The recall covers 2,306 units carrying GTIN 00195278343383, distributed worldwide. The FDA has classified the action as Class II.

The enforcement record describes the reboot condition and reports no injuries. Sites running version V3.0.5 should follow GE HealthCare's instructions.

The recalled product

Product
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube
Manufacturer
Wipro GE Healthcare Private Ltd.
Hazard
  • software-defect
  • reboot-loop
  • loss-of-monitoring

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • GTIN 00195278343383
  • Serial Numbers: STV25225456SA
  • STV25225455SA
  • STV25266464SA
  • STV25235817SA
  • STV25235816SA
  • STV25286682SA
  • STV25225594SA
  • STV25235792SA
  • STV25235774SA
  • STV25225550SA
  • STV25235626SA
  • STV25235625SA
  • STV25184592SA
  • STV25235628SA
  • STV25235631SA
  • STV25184595SA
  • STV25235629SA
  • STV25235632SA
  • STV25235630SA

Distribution

Distribution scope not specified by the agency.