The Recall Desk

FDA (Devices) recall action 97646

Wipro GE Healthcare Private Ltd. recalled 2 products on 2025-10-29

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2025-10-29
Critical
0
Units
2,538

Why these products were recalled

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2025-10-29

    Central stations recalled over reboot loop that halts patient monitoring

    GE HealthCare is recalling 173 Carescape Central Stations running software V3.0.5 because the system can enter a continuous reboot cycle, losing central monitoring of connected patients.

    Product
    GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-10-29

    Central monitoring stations recalled over reboot loop at specific network sizes

    GE HealthCare is recalling 2,306 Carescape Central Stations running software V3.0.5 because the system can enter a continuous reboot cycle, losing central monitoring of connected patients.

    Product
    GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube
    Category
    Distribution
    0 states