Smiths Medical SurgiVet pressure transducer recalled for monitoring failures
Smiths Medical is recalling SurgiVet Advisor Tech Accessories due to manufacturing defects in the pressure transducer that may cause inaccurate, interrupted, or failed pressure monitoring.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall with no reported illnesses or injuries attributed to the defect. The potential for inaccurate or interrupted pressure monitoring in a medical device represents a risk-of-harm scenario, but without reported actual harm, this meets the High severity criteria per the rubric.
Plain-English summary
Smiths Medical ASD, Inc. is recalling the SurgiVet Advisor Tech Accessories (Product Code REF V2714) with affected lot numbers 6000465 and 6000466. The product is distributed worldwide due to a manufacturing defect in the TranStar Disposable Pressure Transducer component.
The defect can manifest as either a leak within the pressure transducer or a defect in the transducer chip. Users may experience inaccurate pressure monitoring, interruption in pressure monitoring, or inability to zero the device as a result of this manufacturing defect.
The recalled product
- Product
- smiths medical SurgiVet Advisor Tech Accessories, Product Code REF V2714
- Manufacturer
- Smiths Medical ASD, Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- UDI/DI N/A
- Lot Numbers: 6000465
- 6000466
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 24 recalled by Smiths Medical ASD, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 24 products in this recall →Related recalls
Same manufacturer · Smiths Medical ASD, Inc.
See all →- SevereCADD Solis VIP Ambulatory Infusion Pump Thermal Damage Recall
FDA (Devices) · 2025-05-14
- SevereCADD-Solis Infusion Pump Thermal Damage Risk Recall
FDA (Devices) · 2025-05-14
- CriticalCADD Solis HSPCA Pump wireless connection alarm interruption recall
FDA (Devices) · 2025-05-14
- CriticalCADD-Solis Ambulatory Infusion Pump False Alarm Malfunction
FDA (Devices) · 2025-05-14
- CriticalCADD-Solis VIP Ambulatory Infusion Pump False Occlusion Alarm Recall
FDA (Devices) · 2025-05-14
Same hazard · measurement error
See all →- HighPhilips Evolution Upgrade 1.5T MRI Stiffness Calculation Error
FDA (Devices) · 2026-05-06
- ModeratePhilips MR 7700 MRE Upgrade MRI Equipment Stiffness Value Error Risk
FDA (Devices) · 2026-05-06
- ModeratePhilips MR 7700 MR Elastography device stiffness value error recall
FDA (Devices) · 2026-05-06
- ModeratePhilips SmartPath to Ingenia Elition X MR Elastography Stiffness Value Errors
FDA (Devices) · 2026-05-06
- ModeratePhilips Achieva 1.5T MRI System MRE Protocol Parameter Error Recall
FDA (Devices) · 2026-05-06