FDA (Devices) recall action 95438

Smiths Medical ASD, Inc. recalled 24 products on 2024-10-30

The FDA (Devices) publishes a recall like this one as 24 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 24 pages.

Products
24
Announced
2024-10-30
Critical
0
Units
1,197,300

Why these products were recalled

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–24 of 24

  • HighFDA (Devices)··2024-10-30

    Smiths Medical Recalls TranStar KIDS KIT Blood Sample Kits

    Smiths Medical ASD, Inc. is recalling 800 units of medex TranStar KIDS KIT closed blood sample kits due to a manufacturing defect that may cause inaccurate pressure monitoring.

    Product
    medex TranStar KIDS KIT 19IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9543
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical Recalls TranStar KIDS KIT Blood Sample Kits

    Smiths Medical ASD, Inc. is recalling 900 units of medex TranStar KIDS KIT 31IN. CLOSED BLOOD SAMPLE KIT due to a manufacturing defect that may cause inaccurate pressure monitoring.

    Product
    medex TranStar KIDS KIT 31IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9544
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical TranStar Pressure Transducers Recalled Due to Manufacturing Defect

    Smiths Medical is recalling 1,000 units of medex TranStar DPT pressure transducers that may monitor inaccurately or interrupt monitoring due to a manufacturing defect. Users may be unable to zero the device.

    Product
    medex TranStar DPT w/Stopcock & 30cc Flush 10/EA, Product Code REF MX950SCNT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar Kids blood sample kit recalled for pressure monitoring failures

    Smiths Medical is recalling approximately 3,280 medex TranStar KIDS KIT blood sample devices due to a manufacturing defect causing pressure monitoring failures.

    Product
    medex TranStar KIDS KIT 7INCH CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9547
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Blood Sample Pressure Kit Recalled for Monitoring Defects

    Smiths Medical recalls the medex TranStar KIDS KIT blood sample kit due to manufacturing defects in the pressure transducer that may cause inaccurate or interrupted pressure monitoring, or prevent proper device calibration.

    Product
    medex TranStar KIDS KIT 10IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9548
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Pressure Monitoring Kit Recalled for Manufacturing Defect and Malfunction

    The medex TranStar 84in Single Monitoring Kit is being recalled due to manufacturing defects in its disposable pressure transducer. The defect may cause inaccurate pressure readings, interruption of monitoring, or inability to zero the device.

    Product
    medex TranStar 84in (213cm) Single Monitoring Kit 10/EA, Product Code REF MX9505T
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    medex TranStar Pressure Transducer Stopcock Recalled for Manufacturing Defect

    Smiths Medical is recalling the medex TranStar Stopcock due to a manufacturing defect that can cause inaccurate pressure monitoring, interrupted pressure monitoring, or inability to zero the device.

    Product
    medex TranStar Stopcock w/Ext. 10/EA, Product Code REF MXG300242
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    medex TranStar Neonatal Monitoring Kit Recalled Due to Pressure Transducer Defect

    Smiths Medical is recalling the medex TranStar Neonatal Monitoring Kit due to manufacturing defects in the pressure transducer that may cause inaccurate readings or monitoring failures.

    Product
    medex TranStar Neonatal Monitoring Kit 10/EA, Product Code REF MX9530T
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar pressure monitoring kit recalled for manufacturing defect

    Smiths Medical ASD is recalling the medex TRANSTAR MONITORING KIT due to a manufacturing defect that may cause inaccurate pressure readings, monitoring interruption, or inability to zero the device. Affected units have Lot Number 4270117.

    Product
    medex TRANSTAR MONITORING KIT 1/EA, Product Code REF MX20103R1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical Arterial Kit recalled for pressure monitoring defect

    Smiths Medical ASD is recalling medex Arterial Kits due to a manufacturing defect in the TranStar Disposable Pressure Transducer that may cause inaccurate readings or device failure.

    Product
    medex Arterial Kit 1/EA, Product Code REF MX7827
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical SurgiVet pressure transducer recalled for monitoring failures

    Smiths Medical is recalling SurgiVet Advisor Tech Accessories due to manufacturing defects in the pressure transducer that may cause inaccurate, interrupted, or failed pressure monitoring.

    Product
    smiths medical SurgiVet Advisor Tech Accessories, Product Code REF V2714
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical TranStar Pressure Transducers Recalled for Monitoring Defect

    Smiths Medical recalls medex TranStar DPT pressure transducers due to a manufacturing defect that may cause inaccurate or interrupted pressure monitoring. The defect stems from a leak or chip malfunction.

    Product
    medex TranStar DPT w/Stopcock & 3cc Flush 10/EA, Product Code REF MX950SCFT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar Kids Blood Sample Kit Pressure Monitoring Defect Recall

    Smiths Medical is recalling medex TranStar Kids blood sample kits due to a manufacturing defect in the pressure transducer that may cause inaccurate or interrupted pressure monitoring.

    Product
    medex TranStar KIDS KIT 34IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9545
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar pressure monitoring sets recalled due to manufacturing defect

    Smiths Medical is recalling 1,640 units of medex TranStar pressure monitoring sets due to manufacturing defects in the pressure transducer that can cause inaccurate pressure readings or monitoring interruptions.

    Product
    medex TranStar MONITORING SET W/SAMPLING PORT 10/EA, Product Code REF MXG300260
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar Pressure Transducer Recalled Due to Manufacturing Defect

    Smiths Medical is recalling 1110 units of the medex TranStar DPT pressure transducer due to a manufacturing defect that may cause inaccurate or interrupted pressure monitoring. The defect stems from a leak or chip failure within the device.

    Product
    medex TranStar DPT w/Two Stopcocks 10/EA, Product Code REF MX950X2SC
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical medex TranStar Pressure Monitoring Kits Recalled for Manufacturing Defect

    Smiths Medical is recalling 5,880 medex TranStar pressure monitoring kits worldwide due to a manufacturing defect that may cause inaccurate readings, monitoring interruption, or inability to zero the device.

    Product
    medex TranStar 60in (152cm) Single Monitoring Kit 10/EA, Product Code REF MX9504T
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Medex TranStar Pressure Transducer Recalled for Manufacturing Defect Affecting Monitoring

    Smiths Medical is recalling 3080 medex TranStar pressure transducers with a manufacturing defect that may cause inaccurate, interrupted, or absent pressure monitoring. The defect stems from leaks or chip failures in the transducer.

    Product
    medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    medex TranStar Pressure Transducer Recall Due to Manufacturing Defect

    Smiths Medical recalls medex TranStar surgical pressure transducers worldwide due to manufacturing defects that may cause inaccurate or interrupted pressure monitoring.

    Product
    medex TranStar Single OR Set 1/EA, Product Code REF MX20846R2
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    medex TranStar Pressure Monitoring Kit Recalled for Manufacturing Defect

    Smiths Medical is recalling the medex TranStar 72-inch Double Monitoring Kit due to a manufacturing defect in the pressure transducer that may cause inaccurate pressure readings or monitoring interruptions. Approximately 6,590 units have been distributed worldwide.

    Product
    medex TranStar 72in (183cm) Double Monitoring Kit 10/EA, Product Code REF MX9502T
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2024-10-30

    Smiths Medical Recalls 50 medex NICU Monitoring Kits Due to Defect

    Smiths Medical ASD, Inc. is recalling 50 medex NICU Monitoring Kits due to a manufacturing defect that may cause inaccurate pressure monitoring.

    Product
    medex NICU Monitoring Kit Set 1/EA, Product Code REF MX20086
    Category
    Distribution
    0 states