Blood Sample Pressure Kit Recalled for Monitoring Defects
Smiths Medical recalls the medex TranStar KIDS KIT blood sample kit due to manufacturing defects in the pressure transducer that may cause inaccurate or interrupted pressure monitoring, or prevent proper device calibration.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a risk-of-harm product where injury has not yet been reported. A defect affecting pressure monitoring in a pediatric blood sample kit represents a potential clinical hazard, though no illnesses or injuries are mentioned in the recall notice.
Plain-English summary
Smiths Medical ASD, Inc. is recalling the medex TranStar KIDS KIT—a 10-inch closed blood sample kit with integrated pressure monitoring designed for pediatric patients. The product includes a 30cc/hr flush device and disposable pressure transducer.
The recall involves 3,490 units with worldwide distribution. A manufacturing defect in the disposable pressure transducer or its internal chip can cause leaks or component failures that compromise device function.
As a result of these defects, users may experience inaccurate pressure readings, interruptions in pressure monitoring during use, or inability to properly zero (calibrate) the device. The affected lot numbers are 4221225, 4250996, 4297929, 4304897, 4316671, 4323138, 4328494, 4387189, and 4404713.
The recalled product
- Product
- medex TranStar KIDS KIT 10IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9548
- Manufacturer
- Smiths Medical ASD, Inc.
- Affected units
- 3,490
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (11)
- UDI/DI 50351688503006 (case)
- 10351688503008 (pouch)
- Lot Numbers: 4221225
- 4250996
- 4297929
- 4304897
- 4316671
- 4323138
- 4328494
- 4387189
- 4404713
Distribution
Distribution scope not specified by the agency.
Part of a larger recall action
This product is one of 24 recalled by Smiths Medical ASD, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 24 products in this recall →Related recalls
Same manufacturer · Smiths Medical ASD, Inc.
See all →- SevereCADD Solis VIP Ambulatory Infusion Pump Thermal Damage Recall
FDA (Devices) · 2025-05-14
- SevereCADD-Solis Infusion Pump Thermal Damage Risk Recall
FDA (Devices) · 2025-05-14
- CriticalCADD-Solis Ambulatory Infusion Pump False Alarm Malfunction
FDA (Devices) · 2025-05-14
- CriticalCADD Solis HSPCA Pump wireless connection alarm interruption recall
FDA (Devices) · 2025-05-14
- CriticalCADD-Solis VIP Ambulatory Infusion Pump False Occlusion Alarm Recall
FDA (Devices) · 2025-05-14
Same hazard · measurement error
See all →- ModeratePhilips MR 7700 MRE Upgrade MRI Equipment Stiffness Value Error Risk
FDA (Devices) · 2026-05-06
- HighPhilips Evolution Upgrade 1.5T MRI Stiffness Calculation Error
FDA (Devices) · 2026-05-06
- ModeratePhilips SmartPath to Ingenia Elition X MR Elastography Stiffness Value Errors
FDA (Devices) · 2026-05-06
- ModeratePhilips Ingenia Ambition X MRI Systems Stiffness Value Error Recall
FDA (Devices) · 2026-05-06
- ModeratePhilips Achieva 1.5T MRI System MRE Protocol Parameter Error Recall
FDA (Devices) · 2026-05-06
Similar recalls
- HighSmiths Medical blood sampling system recalled for pressure reading errors
FDA (Devices) · 2021-10-20
- HighBlood sampling system sets recalled for pressure reading and measurement errors
FDA (Devices) · 2021-10-20
- HighSmiths Medical Antishunt Sets Recalled Due to Pressure and Reading Errors
FDA (Devices) · 2021-10-20
- HighSmiths Medical blood sampling system recalled for pressure measurement errors
FDA (Devices) · 2021-10-20
- HighMedical Device Recall: Blood Sampling System Pressure and Reading Errors
FDA (Devices) · 2021-10-20
- HighSmiths Medical Blood Sampling System Recalled for Pressure Malfunction and Incorrect Readings
FDA (Devices) · 2021-10-20