Brainlab ExacTrac Vero Radiation Therapy System Recall for Positioning Error
Brainlab is recalling ExacTrac Vero systems because a software error may cause the treatment couch to miss the planned position, potentially affecting radiation therapy accuracy.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (radiation therapy positioning) where injury has not yet been reported, fitting the criteria for a High severity score.
Plain-English summary
Brainlab AG is recalling 28 units of the ExacTrac Vero system, a medical device used in conjunction with the MHI-TM2000 Radiation Therapy Linear Accelerator System. The affected units are ExacTrac Vero software versions 3.5(.x), specifically catalog numbers 46228, 46238, 46216, and 46218. The system is intended to analyze patient position and calculate necessary correction shifts for the radiation therapy couch.
The recall was initiated after Brainlab internally detected that under specific conditions, the ExacTrac Vero software may not correctly account for ring angle corrections during the calculation of the couch shift. This error may result in the treatment couch not moving to the exact planned position, creating a deviation between the planned and actual treatment target position. The source notes that this deviation may not be clearly visible to the user if a verification image of the patient position is not acquired after the couch shift is performed.
The affected units were distributed worldwide, including in the United States, Belgium, Canada, France, Germany, Italy, Japan, and South Korea. The source text does not provide specific instructions for consumers or healthcare providers on how to resolve the issue, nor does it report any illnesses or injuries associated with this defect.
The recalled product
- Product
- BRAINLAB EXACTRAC VERO, Model/Catalog Numbers: 46228 EXACTRAC VERO 3.5 46238 EXACTRAC VERO 3.5 46216 EXACTRAC VERO SW UPDATE 3.5.2 TO 3.5.3 46218 EXACTRAC VERO SW UPDATE 3.5.3 TO 3.5.4 The ExacTrac Vero system is intended to be used in conjunction with the MHI-TM2000…
- Manufacturer
- Brainlab AG
- Affected units
- 28
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Brainlab AG
See all →- HighMedical Device Software May Merge Patient Records Without Notification
FDA (Devices) · 2025-03-26
- HighSurgical navigation system recalled for potential mechanical instability
FDA (Devices) · 2023-04-05
- HighRadiation Therapy Software May Not Properly Execute Heart-Protection Breath Hold
FDA (Devices) · 2023-03-22
- HighExacTrac Dynamic Software Yaw Angle Error May Affect Radiotherapy Positioning
FDA (Devices) · 2022-03-16
- HighRadiotherapy positioning software may delay patient movement display
FDA (Devices) · 2021-06-16
Same hazard · positioning error
See all →- SevereHitachi Proton Beam Therapy System Recall Over Positioning Software Issue
FDA (Devices) · 2026-03-04
- HighPatient positioning systems recalled over stereotactic radiosurgery tolerance errors
FDA (Devices) · 2026-01-21
- HighDORO Easy-Connect Navigation Adaptor compatibility issue recall
FDA (Devices) · 2025-06-04
- HighSpinal Fusion Device Recall: Setscrew Insertion and Tightening Issues
FDA (Devices) · 2024-05-15
- HighTruDi Navigation System position discrepancy during ENT surgery
FDA (Devices) · 2023-11-01
Similar recalls
- HighBrainlab ExacTrac Vero 3.5 Patient Positioning System Recalled for Positioning Errors
FDA (Devices) · 2016-06-15
- SevereBrainlab ExacTrac 6.0 Radiation Positioning System Recalled for Targeting Error
FDA (Devices) · 2014-05-07
- HighBrainlab ExacTrac Software Recall for Potential Radiation Misalignment
FDA (Devices) · 2014-08-13
- HighBrainlab ExacTrac 6.0 Patient Positioning System Recalled for Display Error
FDA (Devices) · 2015-09-02
- SevereBrainlab ExacTrac Dynamic Radiation Therapy System Software Recall
FDA (Devices) · 2021-03-17
- HighBrainlab ExacTrac 6.x Software Recalled for Potential Positioning Errors
FDA (Devices) · 2015-05-13