TruDi Navigation System position discrepancy during ENT surgery
A surgical navigation system used in ENT procedures may display incorrect curette tip positions, potentially causing delayed surgery, cerebrospinal fluid leaks, vision loss, or skull damage. The recall affects 141 units across 30 US states.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II recall of a surgical navigation system with a positioning error that could result in serious surgical complications including cerebrospinal fluid leaks, vision loss, and skull base damage. No illnesses or injuries have been reported; the hazard is potential rather than confirmed, which caps the score at 3 per the rubric.
Plain-English summary
Acclarent, Inc. is recalling the TruDi Navigation System (Model FG-2000-00, Catalog ENS022B) when used with TruDi Curette models TDC0005Z and TDC0005. The affected software version is V2.3.1 Update (2.3.1.144 and 2.3.1.166). The recall involves 141 units distributed across 30 US states.
The TruDi Navigation System is used during ear, nose, and throat (ENT) and skull base surgical procedures to help surgeons visualize and track the location of the curette tip. The recalled units contain a discrepancy between the actual location of the curette tip and the location displayed on the navigation system.
This positioning discrepancy may result in delayed or prolonged surgical procedures, cerebrospinal fluid leaks, visual impairment, or damage to skull base structures. These complications can occur during surgical procedures when the surgeon relies on the incorrect position information displayed by the system.
The recalled product
- Product
- TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005
- Manufacturer
- Acclarent, Inc.
- Affected units
- 141
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- TruDi Navigation System
- UDI-DI: 10846835018639
- software version: V2.3.1 Update (2.3.1.144 and 2.3.1.166)
Distribution
Related recalls
Same manufacturer · Acclarent, Inc.
See all →- HighSurgical Suction Instruments Recalled for Calibration Defect Risk
FDA (Devices) · 2021-12-29
- ModerateAcclarent Recalls TruDi NAV Suction Instruments Due to Calibration Issues
FDA (Devices) · 2020-01-29
- ModerateAcclarent Recalls TruDi NAV Suction Instruments Due to Calibration Issues
FDA (Devices) · 2020-01-29
- ModerateAcclarent Recalls TruDi NAV Suction Instruments Due to Calibration Issues
FDA (Devices) · 2020-01-29
- LowAcclarent Recalls Expired RELIEVA FLEX Sinus Guide Catheter Units
FDA (Devices) · 2015-06-17
Same hazard · positioning error
See all →- SevereHitachi Proton Beam Therapy System Recall Over Positioning Software Issue
FDA (Devices) · 2026-03-04
- HighPatient positioning systems recalled over stereotactic radiosurgery tolerance errors
FDA (Devices) · 2026-01-21
- HighDORO Easy-Connect Navigation Adaptor compatibility issue recall
FDA (Devices) · 2025-06-04
- HighSpinal Fusion Device Recall: Setscrew Insertion and Tightening Issues
FDA (Devices) · 2024-05-15
- HighRadiation therapy software update displays incorrect laser positioning
FDA (Devices) · 2022-05-04