Varian Head Frame Posts Recalled Due to Rotation Risk
Varian Medical Systems is recalling 75 Head Frame Posts because they can rotate easily on the mounting cam. No serious injuries have been reported.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device) where injury has not yet been reported, which aligns with the High severity criteria in the rubric.
Plain-English summary
Varian Medical Systems, Inc. is recalling 75 units of the Varian Head Frame System, specifically the Head Frame Posts (PN1008016) and Post Kits. These devices are used with computed tomography scanners and charged particle accelerators for imaging and immobilization during stereotactic radiosurgery or radiotherapy treatments for cranial lesions and tumors.
The recall was initiated after a report indicated that the head frame posts could be easily rotated when attached to the mounting cam on the metal head ring. This mechanical instability poses a risk to patient immobilization during treatment. The source text explicitly states that there was no report of serious injury due to this issue.
The affected devices were distributed in the United States and internationally to one site each in Canada, India, and Morocco. Users of these devices should contact Varian Medical Systems for instructions on how to address the defect.
The recalled product
- Product
- Varian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The Varian Head Frame System is for use with a computed tomography scanner to perform imaging for treatment planning and a charged particle accelerator to perform immobilizatio
- Manufacturer
- Varian Medical Systems, Inc.
- Affected units
- 75
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All Head Frame Posts and Post kits
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Varian Medical Systems, Inc.
See all →- HighVarian Multileaf Collimator Recall Due to Software-Related Leaf Movement Failure
FDA (Devices) · 2020-05-13
- HighVarian Eclipse Treatment Planning System Proton Algorithm Recall
FDA (Devices) · 2019-01-30
- HighVarian Eclipse Treatment Planning System Software Recall for Contouring Errors
FDA (Devices) · 2017-01-18
- HighVarian AlignRT Plus Optical Surface Monitoring System Recall for Parameter Errors
FDA (Devices) · 2016-07-20
- HighVarian Optical Surface Monitoring System Recall for Unintended Couch Parameter Changes
FDA (Devices) · 2016-07-20
Same hazard · mechanical failure
See all →- HighIMRIS ORT300 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighIMRIS ORT200 Operating Room Table Recalled for Mechanical Failure Risk
FDA (Devices) · 2026-09-02
- HighForest River Transit Buses Recalled for Wheelchair Lift Control Failure
NHTSA · 2026-09-01
- HighBlue Bird Recalls Vision School Buses Due to Wheelchair Lift Defect
NHTSA · 2026-08-31
- SevereMagic Tilt Watercraft Trailers Recalled Due to Winch Weld Failure
NHTSA · 2026-08-28