RayStation Radiation Therapy Planning Software Recalled for Dose Calculation Errors
RaySearch is recalling RayStation software versions 2.5 through 5.0 due to dose calculation errors when using specific region of interest settings.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used for radiation therapy planning where dose calculation errors could lead to incorrect treatment, representing a risk of harm where injury has not yet been reported.
Plain-English summary
RaySearch Laboratories AB is recalling RayStation Radiation Therapy Treatment Planning System software versions 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0, and 4.3 (InverseArc 1.0). The recall involves 522 units in the United States and 846 units outside the United States. The software is used to plan and analyze radiation therapy treatments.
The recall was initiated after RaySearch discovered issues with dose calculations when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline. The problem was identified during cooperation with a customer experimenting with dose calculation on a phantom. The affected build numbers include 2.5.0.144, 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.42, 4.3.0.14, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.4.4, 4.7.5.4, 5.0.0.37, 5.0.1.11, and 5.0.2.35.
Healthcare providers and users of the affected software should contact RaySearch Laboratories AB for information on the correction. The source text does not specify a specific action for consumers, as the product is a professional medical device.
The recalled product
- Product
- RayStation 2.5, 3.0, 3.5, 4.0, 4.5, 4.7, 5.0 and 4.3 (InverseArc 1.0) -- Radiation Therapy Treatment Planning System RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Hazard
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Build Numbers: 2.5.0.144
- 2.5.1.89
- 3.0.0.251
- 3.5.0.16
- 3.5.1.6
- 4.0.0.14
- 4.0.1.4
- 4.0.2.9
- 4.0.3.42
- 4.3.0.14
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.4.4
- 4.7.5.4
- 5.0.0.37
- 5.0.1.11 or 5.0.2.35.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · dose calculation error
See all →- HighRadiation Therapy Planning System Density Calculation Inconsistency in RayStation
FDA (Devices) · 2025-02-26
- HighRayStation radiation therapy planning software density uncertainty calculation inconsistency
FDA (Devices) · 2025-02-26
- HighMedical device recall: RayStation radiation therapy software dose calculation inconsistency
FDA (Devices) · 2025-02-26
- HighRayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency
FDA (Devices) · 2025-02-26
- HighRadiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations
FDA (Devices) · 2025-02-26
Similar recalls
- HighRaySearch RayStation Software Recalled for Dose Calculation Errors
FDA (Devices) · 2015-06-03
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Errors
FDA (Devices) · 2016-12-14
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- SevereRaySearch Ray Station Software Recalled for Dose Miscalculations
FDA (Devices) · 2014-04-30
- HighRayStation Software Recall for Radiation Therapy Dose Calculation Errors
FDA (Devices) · 2014-06-25