The Recall Desk

Manufacturer

B Braun Medical Inc

184 recalls in our database name B Braun Medical Inc as the manufacturing or recalling firm.

About this listing

The manufacturer string comes from the agency’s own source notice — for FDA recalls, the “recalling firm” field; for NHTSA recalls, the vehicle manufacturer; for CPSC recalls, the firm-of-record on the joint press release. A single corporate family can appear under multiple manufacturer strings as agency notices aren’t standardized on a canonical name. If you don’t see recalls you expect, try a parent-company search or a broader full-text search.

Severity scores below are computed from the agency’s classification and the recall text, never from manufacturer behaviour or history. Many manufacturers issue routine corrective actions (software updates, labelling corrections) that are correctly scored low — having recalls on this list isn’t inherently a quality signal.

176–184 of 184

  • HighFDA (Devices)·Z-2434-2024·2024-07-31

    Perfusor Space Syringe Pump Recalled Due to Monoject Syringe Incompatibility

    B Braun Medical's Perfusor Space Syringe Pump may not properly recognize Cardinal Health Monoject syringes due to manufacturing changes, risking medication overdose, underdose, or therapy delays.

    Product
    Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2133-2024·2024-06-19

    Stimuplex A Anesthesia Needles Recalled for Undisclosed DEHP in Glue

    B Braun Medical Inc is recalling 108,811 units of Stimuplex A anesthesia needles because the devices are labeled as DEHP-free but the glue contains traces of DEHP.

    Product
    Stimuplex A, 30 DEG, 20GX6", 0.90x150mm, Catalogue Number: 4894278
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2134-2024·2024-06-19

    Stimuplex A Nerve Needles Recalled for Mislabeled DEHP Content

    B Braun Medical Inc is recalling over 1.1 million Stimuplex A nerve block needles labeled as DEHP-free, but the adhesive contains traces of DEHP. The mislabeling poses a risk to users who selected the product specifically to avoid DEHP exposure.

    Product
    Stimuplex A, 30 DEG, 22GX2", 0.70x50mm, Catalogue Number: 4894502
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2131-2024·2024-06-19

    Stimuplex A Injection Needles Recalled for DEHP Mislabeling

    B Braun Medical Inc is recalling Stimuplex A injection needles because the devices are labeled as DEHP-free, but the glue used contains traces of DEHP. The recall affects 20,850 units distributed in the US and Canada.

    Product
    Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2132-2024·2024-06-19

    Medical device nerve block needles incorrectly labeled as DEHP-free

    Stimuplex A nerve block needles labeled as DEHP-free contain trace amounts of DEHP in the adhesive. B Braun is recalling approximately 1.78 million units distributed in the US and Canada.

    Product
    Stimuplex A, 30 DEG, 21GX4", 0.80x100mm, Catalogue Number: 4894260
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0009-2023·2022-10-12

    Introcan Safety Catheter Recalled for Potential Leakage at Hub

    B. Braun is recalling 94,000 Introcan Safety FEP 14G catheters due to potential leakage at the catheter hub. Affected units were distributed in the US and Canada.

    Product
    Introcan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arterial access.
    Category
    Medical Device
    Distribution
    15 states
  • HighFDA (Devices)·Z-0873-2022·2022-04-20

    Hemodialysis Conductivity Sensors recalled for potential cracks and dialysate leaks

    B Braun Medical Inc is recalling conductivity sensors used in Dialog hemodialysis systems due to potential cracking during use, which could cause leaks in the dialysate circuit and improper ultrafiltration. Affected patients should contact their healthcare provider.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 0
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0872-2022·2022-04-20

    Conductivity Sensors in Dialog Hemodialysis Systems May Crack and Leak

    B Braun Medical Inc is recalling conductivity sensors used in Dialog and Dialog+ hemodialysis systems due to potential cracks that could cause dialysate circuit leaks, affecting ultrafiltration during treatment.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 0
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0819-2022·2022-03-30

    Infusion Pump Risk of Malfunction in B. Braun Perfusor Space Devices

    B. Braun is recalling 6 Perfusor Space infusion pump units that may not function properly due to inadequate inspection, potentially causing incorrect medication dosing and therapy delays.

    Product
    PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made solutions Catalog Number: 8713030U
    Category
    Medical Device
    Distribution
    2 states