Sterilization Error recalls

91 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
91
Critical
0
Severe
0
Most recent
2026-08-26

Of the 91 sterilization error recalls we have scored against our rubric, 0 (0%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 0 (0%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all sterilization error recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

51–75 of 91

  • ModerateFDA (Devices)·Z-0664-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Obturator Packages Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of 11mm Obturator packages because the instructions for use do not support specific sterilization methods.

    Product
    PKG, OBTURATOR, 11MM, FOR HASSON CANNULA, P/N 0250080158. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) p
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0625-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the provided instructions for use do not support specific sterilization methods.

    Product
    PKG, AXIAL HANDLE, W/RATCHET, INSULATED SHAFT, P/N 0250080235. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0757-2015·2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM PEEK HANDLE, 45CM, P/N 0250181164 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0630-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Grasper Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic grasper forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, NARROW JAW, P/N 0250080047 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0792-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Laparoscopic Graspers Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 laparoscopic atraumatic graspers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, ATRAUMATIC GRASPER, DOUBLE ACTION, 20CM, P/N 0250282100 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fr
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0784-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of 3mm Bullet Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BULLET FORCEPS, 29CM, P/N 0250282057 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0599-2015·2014-12-24

    Stryker Laparoscopic Knot Pushers Recalled for Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic knot pushers because the instructions for use do not support specific sterilization methods.

    Product
    PKG, KNOT PUSHER 5MM OPEN, P/N 0250080132 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0694-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of bipolar forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080360. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) por
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0717-2015·2014-12-24

    Stryker Laparoscopic Instruments Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, ALLIS DOUBLE ROW TEETH, P/N 0250080592 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allo
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0601-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Palpation Probes Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 laparoscopic palpation probes because the instructions for use do not support specific sterilization methods.

    Product
    PKG, PALPATION PROBE GRADUATED 5MM, P/N 0250080135; Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0675-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BLUNT FORCEPS, BLUNT, SERRATED, SPOON, P/N 0250080313. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0654-2015·2014-12-24

    Stryker Endoscopy Recalls Tissue Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of tissue grasper forceps because the instructions for use do not support two specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, TISSUE, P/N 0250080284. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which a
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0704-2015·2014-12-24

    Stryker Endoscopy Recalls Maryland Dissecting Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Maryland Dissecting Forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, MARYLAND DISSECTING FORCEPS, CURVED LEFT, 45CM, P/N 00250080578 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0725-2015·2014-12-24

    Stryker Laparoscopic Graspers Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, RETENTION GRASPER, 33CM, P/N 0250080697 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) por
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0795-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM X 20CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250282104 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fre
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0734-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, PADDLE BABCOCK, 45CM, P/N 0250080727 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0706-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, DISSECTING & GRASPER FORCEPS, SERRATED, 45CM, P/N 0250080579 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have rad
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0787-2015·2014-12-24

    Stryker Endoscopy Recalls Biopsy Punch Instruments Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 biopsy punch instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, BIOPSY PUNCH, 29CM, P/N 0250282060 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports wh
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0620-2015·2014-12-24

    Stryker Maryland Forceps Recalled Due to Sterilization Parameter Errors

    Stryker Endoscopy is recalling Maryland Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, MARYLAND FORCEPS, CURVED, LEFT, P/N 0250080042. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0779-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, FUNDUS GRASPER, 29CPKG, 3MM INSERT, ALLIGATOR GRASPER, 29CM, P/N 0250282051 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instr
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0752-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissecting forceps units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, DISSECTING FORCEPS, P/N 0250080762 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) p
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0798-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Instruments Due to Sterilization Labeling Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, CANNULA 8MM AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080172 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio freq
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0711-2015·2014-12-24

    Stryker Endoscopy Recalls Babcock Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of Babcock Grasper Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, BABCOCK GRASPER FORCEPS, 45CM, P/N 0250080586 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (R
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0751-2015·2014-12-24

    Stryker Laparoscopic Dissector Recall Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissectors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250080761 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio fre
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0607-2015·2014-12-24

    Stryker Endoscopy Recalls Hasson Cannula Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Hasson Cannula units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, HASSON CANNULA, 5.5MM, AUTOMATIC VALVE, STOP COCK, P/N 0250080176. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio
    Category
    Medical Device
    Distribution
    Distributed nationwide