Stryker Endoscopy Recalls Hasson Cannula Due to Sterilization Instructions

Stryker Endoscopy is recalling 180,573 Hasson Cannula units because the instructions for use do not support specific sterilization methods.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves documentation errors regarding sterilization methods without reported injuries or illnesses, fitting the criteria for a moderate severity score.

Plain-English summary

Stryker Endoscopy is recalling 180,573 units of the PKG, HASSON CANNULA, 5.5MM, AUTOMATIC VALVE, STOP COCK (P/N 0250080176). The recall covers units distributed worldwide, including the United States and numerous international locations.

The recall was initiated because the parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. The device is intended for cutting, holding, grasping, and manipulating tissue during laparoscopic procedures.

Healthcare facilities and distributors should stop using the affected devices and contact Stryker Endoscopy for further instructions regarding the recall.

The recalled product

Product
PKG, HASSON CANNULA, 5.5MM, AUTOMATIC VALVE, STOP COCK, P/N 0250080176. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio
Manufacturer
Stryker Endoscopy
Affected units
180,573

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • 29583 34963

Distribution

Distributed nationwide across the United States.

The notice also names 2 states specifically:

Part of a larger recall action

This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 207 products in this recall →