The Recall Desk

Hazard

Software Defect recalls

1,375 recalls in our database tag this hazard, drawn from FDA, USDA FSIS, CPSC, and NHTSA source notices.

What the numbers show

Total
1,375
Critical
14
Severe
106
Most recent
2026-10-21

Of the 1,375 software defect recalls we have scored against our rubric, 14 (1%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 106 (8%) are Severe.

About this hazard tag

Hazard tags are short, machine-readable labels we attach to every recall to make cross-cutting queries possible — “all software defect recalls regardless of agency”, “all listeria recalls in California”, “all airbag recalls in 2025”. The tags are extracted from each recall’s source text during the LLM rewrite step and validated against a fixed schema.

A recall can carry multiple tags — a Listeria recall might also be tagged ready-to-eat and undeclared-allergen. Use the full-text search if you want to combine hazard tags with brand or category filters.

226–250 of 1375

  • HighFDA (Devices)·Z-2616-2024·2024-08-21

    BioButton Remote Monitoring Device Software Accuracy Issue

    A software anomaly causes the BioButton remote monitoring wearable device to report skin temperature readings slightly out of specification, affecting accurate health data collection in home and healthcare settings.

    Product
    BioButton REF BIOST06040 with Software version 3.0.20 The BioButton System is a remote monitoring wearable device intended for continuous collection of physiological data in home and healthcare settings while the patient is at rest.
    Category
    Medical Device
    Distribution
    5 states
  • SevereFDA (Devices)·Z-2419-2024·2024-08-14

    CADD-Solis Ambulatory Infusion Pumps with Software Defects Affecting Device Responsiveness

    CADD-Solis Model 2100 infusion pumps with pre-v4.3 software have defects affecting occlusion detection and device responsiveness. Manufacturer recommends updating to software version 4.3 (2023) to resolve the issues.

    Product
    CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: Intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site, epidural space or subarachnoid space. Therapies that…
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2483-2024·2024-08-14

    MRI System Software Error Can Overheat Patient Implants

    GE Medical Systems recalled three SIGNA 7.0T MRI systems due to a software error that can exceed preset safety limits during imaging, potentially overheating implants in patients under certain conditions.

    Product
    SIGNA 7.0T, with affected software versions: 7T29.1, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2491-2024·2024-08-14

    BD PYXIS MEDBANK Cabinet Software Causes Incorrect Medication Labels

    FDA recall of BD PYXIS MEDBANK automated medication dispensing cabinets due to a software defect that prints incorrect bin information on restock labels, potentially leading to medication identification errors.

    Product
    BD PYXIS MEDBANK TWR MN CR-12HH-1FM-P, REF: 169-85 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2492-2024·2024-08-14

    Automated Medication Cabinet Software Bug May Print Wrong Bin Labels

    CareFusion's BD PYXIS MEDBANK medication dispensing cabinets (software 3.9.1.9) have a software defect that may cause restock labels to print incorrect medication bin locations, creating a risk of medication dispensing errors.

    Product
    BD PYXIS MEDBANK TWR MN CR-14HH-P, REF: 169-86 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2474-2024·2024-08-14

    SIGNA MR380 MRI system software may exceed implant heating limits

    GE SIGNA MR380 MRI systems with software versions SV25.5 or SV25.6 may exceed radiofrequency heating limits during certain imaging sequences, potentially causing MR-conditional implants to overheat. The issue occurs in Low SAR Mode when specific T2 FLAIR sequences are used with the Optimized T2 FLAIR option disabled.

    Product
    SIGNA MR380, with affected software versions: SV25.5, SV25.6; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2503-2024·2024-08-14

    BD PYXIS Automated Dispensing Cabinet Software Issue Causes Incorrect Medication Bin Labels

    A software issue in BD PYXIS automated dispensing cabinets may cause incorrect medication bin labels to print during restocking, potentially leading to medication dispensing errors. CareFusion is recalling 16 units.

    Product
    BD PYXIS MEDBANK TWR MN CR-8HH-3FM-P, REF: 169-99 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2500-2024·2024-08-14

    BD PYXIS MEDBANK Software Issue Causes Incorrect Medication Bin Labels

    A software issue in BD PYXIS MEDBANK automated dispensing cabinets can cause incorrect medication bin labels to print during restocking. This could lead to medication dispensing errors in healthcare facilities.

    Product
    BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2501-2024·2024-08-14

    Medication Dispensing Cabinet Software Error Can Print Incorrect Bin Labels

    A software error in CareFusion's BD PYXIS automated medication dispensing cabinet can cause incorrect medication bin numbers to be printed on restock labels, risking medication dispensing errors.

    Product
    BD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P, REF: 169-97 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2490-2024·2024-08-14

    BD PYXIS automated medication cabinet software defect causes incorrect bin labels

    A software defect in BD PYXIS MEDBANK automated medication dispensing cabinets may cause the incorrect medication bin to be printed on restock labels, potentially resulting in medication errors.

    Product
    BD PYXIS MEDBANK TWR MN CR-12HH-1FH, REF: 169-84 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2502-2024·2024-08-14

    Automated Dispensing Cabinet Software Issue Causing Incorrect Medication Bin Labels

    A software defect in the BD PYXIS MEDBANK automated dispensing cabinet causes incorrect medication bin locations to be printed on restock labels, potentially resulting in medication dispensing errors. No illnesses or injuries have been reported.

    Product
    BD PYXIS MEDBANK TWR MN CR-8HH-3FH-P, REF: 169-98 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2472-2024·2024-08-14

    GE SIGNA Creator MRI systems software defect risks patient implant overheating

    GE HealthCare recalled certain SIGNA Creator MRI systems due to a software defect where the predicted magnetic field strength can exceed prescribed limits in Low SAR Mode with specific imaging sequences, potentially causing overheating of MR-conditional implants.

    Product
    SIGNA Creator, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2468-2024·2024-08-14

    GE MR450w 1.5T MRI software defect may cause implant overheating

    GE MR450w 1.5T MRI systems may overheat MR-conditional implants when using specific imaging sequences with Low SAR Mode. No injuries reported; 1,607 units affected worldwide.

    Product
    Optima MR450w 1.5T, with affected software versions: DV24.0, DV25.0, DV25.1, DV26.0, DV29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2495-2024·2024-08-14

    BD PYXIS Medication Cabinet Software Defect Causes Incorrect Restock Labels

    A software issue in BD PYXIS automated medication dispensing cabinets can cause restock labels to print with incorrect bin numbers. Four units are affected nationwide.

    Product
    BD PYXIS MEDBANK TWR MN CR-4HH-1FH-4FM-P, REF: 169-90 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2494-2024·2024-08-14

    Medication Dispensing Cabinet Software Defect Causes Incorrect Bin Labels

    BD PYXIS automated medication dispensing cabinets contain a software issue that may print incorrect bin locations on restock labels. This could result in medications being placed in the wrong bin during restocking.

    Product
    BD PYXIS MEDBANK TWR MN CR-2HH-6FM, REF: 169-89 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2499-2024·2024-08-14

    Automated Medication Dispenser Software Bug Causes Incorrect Bin Labels

    BD PYXIS MEDBANK automated medication dispensers may print incorrect medication bin labels due to a software defect, risking medication handling errors in healthcare settings.

    Product
    BD PYXIS MEDBANK TWR MN CR-6HH-4FM-P, REF: 169-94 containing software version 3.9.1.9
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·24V560000·2024-07-25

    Nissan Ariya inverter software may shut down the EV system

    Nissan is recalling certain 2023 Ariya vehicles because the inverter software may misdiagnose an error and shut down the EV system, resulting in a loss of drive power.

    Product
    NISSAN — NISSAN ARIYA
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2412-2024·2024-07-24

    Surgical Image Guidance Software May Cause Incorrect Toric IOL Axis Placement

    Surgical image guidance software in NGENUITY 3D systems may miscalculate lens axis placement during eye rotation, potentially causing astigmatic error and decreased visual acuity.

    Product
    Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2262-2024·2024-07-10

    BD Alaris System Manager Software Connectivity Issue Recalled Nationwide

    A software issue in BD Alaris System Manager version 12.5.1 may prevent the PC unit from connecting to the server, potentially affecting wireless data transmission. CareFusion 303, Inc. is recalling the affected software version nationwide.

    Product
    BD Alaris System Manager REF: 9601, compatible with BD Alaris PC Units
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·24V459000·2024-06-20

    Cadillac XT5 and XT6 instrument panels may not illuminate signal reminders

    GM is recalling certain 2023-2025 Cadillac XT5 and XT6 vehicles because the instrument panel may not illuminate when hazard warning lights, high beams or turn signals are activated.

    Product
    CADILLAC — CADILLAC XT5, XT6
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighNHTSA·24V436000·2024-06-13

    Radio software may disable rearview camera in multiple Chrysler vehicle models

    Certain 2021-2023 Chrysler, Jeep, Ram, and Dodge vehicles have defective radio software that may prevent the rearview camera from displaying, reducing driver visibility when backing up.

    Product
    JEEP — JEEP, CHRYSLER, RAM GRAND CHEROKEE, VOYAGER, GRAND WAGONEER, COMPASS
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2021-2024·2024-06-12

    GlideScope Core 15-inch Monitor Recalled for Potential Image Loss Software Issue

    Verathon is recalling GlideScope Core 15-inch monitors due to software issues causing potential image loss or degradation. Approximately 2,568 units were distributed across the US and internationally.

    Product
    Brand Name: GlideScope Core 15-inch Monitor Model/Catalog Number: 0570-0404 Product Description: Serialized Monitor
    Category
    Medical Device
    Distribution
    49 states