FDA (Devices) recall action 97816

B Braun Medical Inc recalled 116 products on 2025-12-03

The FDA (Devices) publishes a recall like this one as 116 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 116 pages.

Products
116
Announced
2025-12-03
Critical
0
Units
93,637,186

Why these products were recalled

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

51–75 of 116

  • HighFDA (Devices)··2025-12-03

    IV administration sets recalled over backflow from piggyback containers

    B. Braun Medical is recalling 2,544 IV administration sets over the potential for medication to flow back from secondary piggyback containers into primary containers, and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 124 IN.; Catal
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Large volume IV sets recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 45,288 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 130 IN. IV ADMIN SET W/ 9 I
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Blood administration sets recalled over backflow and priming failure

    B. Braun Medical is recalling 164,568 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, HAND P
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Infusion sets recalled over backflow from secondary bags into primary

    B. Braun Medical is recalling 122,232 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 129 IN, 3 CAR
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Secondary IV sets recalled over backflow into the primary container

    B. Braun Medical is recalling 25,728 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, QUAD MANIFOLD
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Small volume IV sets recalled over backflow and priming failure

    B. Braun Medical is recalling 4,488 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. AESTHESIA IV SET, 127 in.;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Caresite extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 32,500 Caresite extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Caresite Extension Set_Export; Cata
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Outlook pump sets recalled over backflow into the primary container

    B. Braun Medical is recalling 2,352 Outlook pump sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET 3 CARESITE LADS 133
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia IV sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 14,712 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV AD. SET,15 DROPS/ML, W/EXT S
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Outlook infusion sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 6,672 Outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK PUMP SET,3 CARESITE LADS,115
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Standard IV sets recalled over medication backflow from piggyback containers

    B. Braun Medical is recalling 41,976 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia infusion sets recalled over backflow into the primary bag

    B. Braun Medical is recalling 19,392 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15 DROPS/ML 125 I
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    ADDitIV set 352062 recalled over backflow and inability to prime

    B. Braun Medical is recalling 12,700 ADDitIV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. PB602VSL NON-VENT SET 103-IN-DR; Catalog
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Administration set 354209 recalled over backflow and inability to prime

    B. Braun Medical is recalling 89,552 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ANES 15DROP W/2 CARESITE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set NF1270A recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 13,870 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. NF1270A SAFELINE ANESTHESIA
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    SafeDAY set 352646 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 554,015 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM SET 15 DR W/2 SAFEDAY & STOPCO
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    SAFELINE set 352604 recalled over backflow from secondary containers

    B. Braun Medical is recalling 46,250 SAFELINE sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SAFELINE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. UNIVERSAL IV ADMIN SET; Catalog Numb
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set 473309 recalled over backflow and inability to prime

    B. Braun Medical is recalling 11,650 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. AS152V2DLL; Catalog Number: 473309.
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490239 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 5,100 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 103 IN.; Catal
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set 490234 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 27,480 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET ULTRAPORT STOPCOCK, 37IN.;
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    CARESAFE set 354302 recalled over backflow and inability to prime

    B. Braun Medical is recalling 4,438 CARESAFE sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    CARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. CARESAFE IV ADMIN SET w/AIRSTOP, 104 IN
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490349 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 24,648 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    B.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 60 DROPS/ML,141 in.; Cata
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490241 recalled over backflow and inability to prime

    B. Braun Medical is recalling 66,792 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET HF STOPCOCK,10
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490367 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 6,768 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/60 DR/ML, CA
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490242 recalled over backflow into the primary bag

    B. Braun Medical is recalling 129,480 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP/ML, 10
    Category
    Distribution
    0 states