FDA (Devices) recall action 97816

B Braun Medical Inc recalled 116 products on 2025-12-03

The FDA (Devices) publishes a recall like this one as 116 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 116 pages.

Products
116
Announced
2025-12-03
Critical
0
Units
93,637,186

Why these products were recalled

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

76–100 of 116

  • HighFDA (Devices)··2025-12-03

    IV set 490303 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 32,304 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,3 CARESITE LAD
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Blood set 490530 recalled over backflow and inability to prime

    B. Braun Medical is recalling 33,528 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, 3 CARS
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 352636 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 20,694,734 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DR W/2 SAFE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490025 recalled over backflow into the primary container

    B. Braun Medical is recalling 15,250 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. PB1500 15 DROP IV SET W/2 S
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set 456521 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 61,423 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. ANESTHESIA SET W/ SAFEPORT MANIFOL
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set 490238 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 11,800 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ 6 B/C VALVES, 17 IN; Cat
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set V5405 recalled over backflow and inability to prime

    B. Braun Medical is recalling 300 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT/2INJ/FLBL/FML LRLK/CV 34; Catal
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 354210 recalled over backflow and inability to prime

    B. Braun Medical is recalling 1,861,111 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ANES 15DROP W/2 CARESITE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 477002 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 22,928 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 132
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490427 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 224,856 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15 DROPS/ML
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Extension set 456520 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 2,100 units extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. ANES SET W/DUAL FEM.LL SAFEPORT MAN
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia set 490476 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 19,320 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 2-HIGH FLOW STPCK
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia set 490477 recalled over backflow and inability to prime

    B. Braun Medical is recalling 36,552 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 134
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490343 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 1,272 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET,2 CARESITE, TR
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490309 recalled over backflow into the primary bag

    B. Braun Medical is recalling 259,392 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 CARESITE,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490312 recalled over medication backflow between IV containers

    B. Braun Medical is recalling 226,632 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMN SET W/15 DROP, ULTR
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Outlook set 354212 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 162,648 units outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DROP W/2 CARESITE; C
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 480231 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 85,176 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Outlook set 354214 recalled over backflow and inability to prime

    B. Braun Medical is recalling Outlook set catalog number 354214 over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DR W/2 CARESITE FILT
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-12-03

    Bronchoscopy robots recalled because the user manual was password protected

    Noah Medical is recalling 47 Galaxy Systems because users did not have access to the Instructions for Use and User Manual, which was password protected.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490400 recalled over backflow and inability to prime

    B. Braun Medical is recalling 19,104 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/CARESITE EXT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Blood set 480233 recalled over backflow from piggyback containers

    B. Braun Medical is recalling 5,448 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET W/ 170¿
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490357 recalled over backflow from secondary piggyback containers

    B. Braun Medical is recalling 98,688 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 10 DROPS/ML
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Blood set 490355 recalled over backflow into the primary container

    B. Braun Medical is recalling 24,624 units blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET W/ 170
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Blood set 490196 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 7,344 blood administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET 124 IN.
    Category
    Distribution
    0 states