FDA (Devices) recall action 97816

B Braun Medical Inc recalled 116 products on 2025-12-03

The FDA (Devices) publishes a recall like this one as 116 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 116 pages.

Products
116
Announced
2025-12-03
Critical
0
Units
93,637,186

Why these products were recalled

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–50 of 116

  • HighFDA (Devices)··2025-12-03

    Smallbore extension sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 30,250 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. ET08US EXTENSION SET W/ULTRA Y SITE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Gravity IV sets recalled over medication backflow and inability to prime

    B. Braun Medical is recalling 12,200 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. NF1503 15 DR UNIV SPIKE, IN
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    ULTRASITE set 352049-C recalled over backflow and inability to prime

    B. Braun Medical is recalling 41,400 ULTRASITE sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ULTRASITE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 ULTRASITE VLV, 10
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490193 recalled over backflow and inability to prime

    B. Braun Medical is recalling 111,312 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR HF 4-WAY
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 354208 recalled over medication backflow and priming failure

    B. Braun Medical is recalling 126,050 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 15DR, CARESIT
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490488 recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 187,464 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490274 recalled over backflow into the primary container

    B. Braun Medical is recalling 98,568 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 3 CARESITE LAD
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2025-12-03

    Liquid biopsy reports recalled for omitting the companion diagnostic claims page

    Foundation Medicine is recalling 18 FoundationOne Liquid CDx reports that were sent to customers without the companion diagnostic Claims Page.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    SafeDAY set 352635 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 200,917 SafeDAY sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 60 DROP W/2 SAFEDAY; Cat
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490507 recalled over backflow and inability to prime

    B. Braun Medical is recalling 48,672 units iv administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMINISTRATION SET WITH
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Outlook set 354213 recalled over backflow into the primary container

    B. Braun Medical is recalling 18,072 units outlook sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DROP W/3 CARESITE; C
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Short extension sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 13,600 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. EXT SET W/ B/C VLV, ULTRAPORT, 57 I
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia set 490472 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 10,536 units anesthesia iv sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADM 15D 2 CARESITE, SAFE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV extension sets recalled over backflow between piggyback and primary containers

    B. Braun Medical is recalling 43,452 extension sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Numb
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    ADDitIV sets recalled over backflow between primary and secondary containers

    B. Braun Medical is recalling 100 ADDitIV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. V1782F UNIV IV SET W/0.2 MICRON FILTER; C
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490489 recalled over piggyback backflow and inability to prime

    B. Braun Medical is recalling 3,120 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15DR, 92 IN W/
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia set 490468 recalled over backflow and inability to prime

    B. Braun Medical is recalling 58,752 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 117 IN.; Catalog N
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 490504 recalled over piggyback backflow and priming occlusion

    B. Braun Medical is recalling 36,936 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Anesthesia set 490189 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 32,436 anesthesia IV sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/15 DROPS/ML, 108
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Rate flow sets recalled over backflow and inability to prime

    B. Braun Medical is recalling 150,664 RATE FLOW sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    RATE FLOW set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. RATEFLOW ADMIN 15 DROP W/2 CARESITE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    High volume IV sets recalled over piggyback backflow and occlusion

    B. Braun Medical is recalling 190,625 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W/ CAR
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Single lot IV sets recalled over backflow and priming occlusion

    B. Braun Medical is recalling 48 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN 15D, 2 B/C VLV, 3
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Caresite set 354207 recalled over backflow and priming occlusion

    B. Braun Medical is recalling 381,850 Caresite primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Primary Gravity Filtered Set with 2 CARE
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    Primary gravity sets recalled over backflow from secondary containers

    B. Braun Medical is recalling 43,900 ADDitIV primary gravity sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set, 2 Non-Needle-Fre
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2025-12-03

    IV set 457506 recalled over piggyback backflow and occlusion risk

    B. Braun Medical is recalling 1,019 IV administration sets over the potential for medication backflow from secondary piggyback containers and an inability to prime.

    Product
    IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV SET W/2 CARESITE & 1 ULT
    Category
    Distribution
    0 states