The Recall Desk

FDA (Devices) recall action 92328

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalled 120 products on 2023-06-28

The FDA (Devices) publishes a recall like this one as 120 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 120 pages.

Products
120
Announced
2023-06-28
Critical
6
Units
810,580

Why these products were recalled

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

51–75 of 120

  • CriticalFDA (Devices)··2023-06-28

    Medtronic MIRRO MRI Implantable Defibrillator Recall for Potential Therapy Failure Risk

    Medtronic's MIRRO MRI implantable cardioverter defibrillators with a specific glassed feedthrough may rarely fail to deliver adequate energy output during high-voltage therapy, affecting approximately 2,831 units distributed worldwide.

    Product
    ICD-DR DDME3D1 MIRRO MRI, Model Number DDME3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Deliver Reduced or No Energy During Therapy

    Certain Medtronic implantable cardioverter defibrillators with a specific feedthrough may produce reduced or no energy output during high voltage therapy. The rare defect affects 16 units distributed nationwide.

    Product
    ICD-VR DVBB1D1 EVERA XT IS1/DF1 US, Model Number DVBB1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Critical Heart Therapy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy delivery due to a defective feedthrough component. No injuries have been reported, but patients may not receive proper treatment for life-threatening heart rhythms.

    Product
    CRTD COBALT XT HF QUAD MRI IS4 DF4, Model Number DTPA2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Risk Failing to Deliver Electrical Therapy

    Medtronic is recalling 404 units of implantable cardioverter defibrillators and cardiac resynchronization devices due to a rare risk of reduced or absent energy output during therapy. The defect involves a specific manufacturing feedthrough component.

    Product
    CRTD CROME HF QUAD MRI IS4 DF1, Model Number DTPC2Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Risk Failure to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) with a specific feedthrough may fail to deliver high-voltage therapy. The FDA is recalling 28,172 units as a Class I safety issue.

    Product
    CRTD DTMB2QQ AMPLIA MRI QUAD OUS DF4, Model Number DTMB2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic ICD-VR PRIMO MRI Defibrillators: Potential Therapy Delivery Failure

    Certain Medtronic implantable cardioverter defibrillators and CRT-Ds may fail to deliver adequate energy during high-voltage therapy due to a feedthrough component defect. This affects 5,215 units distributed worldwide.

    Product
    ICD-VR DVMD3D1 PRIMO MRI, Model Number DVMD3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Cardiac Defibrillators Potentially Fail to Deliver High-Voltage Therapy

    Medtronic's implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) may fail to deliver high-voltage therapy due to a defect in specific glassed feedthrough components. The FDA classified this as a Class I recall due to potential for serious harm.

    Product
    CRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Medtronic issued a Class I recall for certain implantable cardioverter defibrillators due to a rare risk that high-voltage therapy may not deliver. Affected models include DTMA1QQ devices made with a specific feedthrough component.

    Product
    CRTD DTMA1QQ CLARIA MRI QUAD US DF4, Model Number DTMA1QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic CRT-D BRAVA Implantable Defibrillators—Potential High-Voltage Therapy Failure

    Medtronic implantable cardioverter defibrillators with specific feedthrough components may fail to deliver high-voltage therapy during treatment. Affected units are being recalled due to risk of reduced or absent energy output.

    Product
    CRT-D DTBC2QQ BRAVA QUAD IS4/DF4 INTL, Model Number DTBC2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardiac Defibrillators May Fail During High-Voltage Therapy

    Certain Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy due to a defective feedthrough component. Affected devices were distributed nationwide and worldwide.

    Product
    CRTD DTMA2Q1 CLARIA MRI QUAD OUS DF1, Model Number DTMA2Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Recalled: Potential Therapy Energy Delivery Failure

    Medtronic is recalling implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) that may deliver reduced or no energy during high-voltage therapy. The recall affects approximately 6,536 units with a specific feedthrough component.

    Product
    CRTD DTMB2D1 AMPLIA MRI OUS DF1, Model Number DTMB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-28

    Implantable Defibrillators Recalled for Potential High-Voltage Therapy Failure

    Medtronic CRT-D and ICD devices with a specific feedthrough may fail to deliver high-voltage therapy. The FDA Class I recall affects multiple lot numbers due to rare potential for reduced or no energy output during critical treatment.

    Product
    CRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable cardiac defibrillators recalled for potential therapy failure risk

    Medtronic recalls ICD COBALT VR MRI DF4 defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy. The defect affects devices manufactured with a specific glassed feedthrough component.

    Product
    ICD COBALT VR MRI DF4, Model Number DVPB3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators May Fail Therapy Energy Delivery

    Medtronic recalls 3 COBALT XT implantable defibrillators that may deliver reduced or no energy during therapy. A defect in the device feedthrough could prevent effective treatment of abnormal heart rhythms.

    Product
    CRTD DTPA2D1PX COBALT XT HF QUAD OUS, Model Number DTPA2D1PX; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Failure During High-Voltage Treatment

    Medtronic is recalling specific Implantable Cardioverter Defibrillator (ICD) models due to a rare potential for reduced or no energy output during high-voltage therapy. The issue is related to a specific manufacturing defect in the device feedthrough.

    Product
    ICD EVERA XT DR GOLD CTD, Model Number DDBB1D4G; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    ICD defibrillators risk reduced or absent therapy output from feedthrough defect

    Certain ICD defibrillators may produce reduced or no energy output during high-voltage therapy due to a feedthrough defect. FDA Class I recall affects approximately 2,031 units nationwide and worldwide.

    Product
    ICD CROME VR MRI DF4, Model Number DVPC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling approximately 10,900 ICD and CRT-D devices that may fail to deliver high-voltage therapy due to a feedthrough component defect. No illnesses reported, but the rare defect poses risk of inadequate device function during critical treatment.

    Product
    ICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable cardioverter defibrillators recalled for potential therapy delivery failure

    Medtronic implantable cardioverter defibrillators (ICDs) with a specific glassed feedthrough are recalled due to a rare potential for reduced or no energy output during high voltage therapy.

    Product
    ICD DDPA2D1G COBALT XT DR MRI DF1 GOLD, Model Number DDPA2D1G; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable defibrillators may fail to deliver high-voltage therapy during treatment

    Certain Medtronic ICD and CRT-D devices with a specific feedthrough may fail to deliver energy during high-voltage therapy. This rare defect affects 9,325 units worldwide and could compromise cardiac protection.

    Product
    ICD COBALT XT VR MRI DF4, Model Number DVPA2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Energy Output Failure

    Medtronic Inc. is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) due to a rare potential for reduced or no-energy output during high voltage therapy. Affected devices contain a specific glassed feedthrough.

    Product
    CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic implantable defibrillators with a specific feedthrough may fail to deliver adequate energy during high voltage therapy. Approximately 476 affected units distributed nationwide and worldwide.

    Product
    CRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    3,570 implantable cardioverter defibrillators and cardiac resynchronization defibrillators with a specific feedthrough defect may fail to deliver high-voltage therapy when needed.

    Product
    CRTD DTMA2D1 CLARIA MRI OUS DF1, Model Number DTMA2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Produce Reduced or No Energy Output

    Medtronic implantable cardioverter defibrillators may produce reduced or no-energy output during high voltage therapy due to a defective feedthrough component. The defect affects 176 units distributed worldwide.

    Product
    ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic ICD COBALT XT implants may deliver reduced or no therapy

    Medtronic's ICD COBALT XT HF QUAD implantable defibrillators may deliver reduced or no energy output during high voltage therapy. The FDA has classified this as a Class I recall.

    Product
    ICD COBALT XT HF QUAD OUS, Model Number DTPA2QQG; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling 1,754 CRT-D and ICD implantable defibrillators with a specific feedthrough component that may not deliver therapy when needed. Devices were distributed nationwide and worldwide.

    Product
    CRT-D DTBC2Q1 BRAVA QUAD IS4/DF1 INTL, Model Number DTBC2Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states