The Recall Desk

FDA (Devices) recall action 92328

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalled 120 products on 2023-06-28

The FDA (Devices) publishes a recall like this one as 120 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 120 pages.

Products
120
Announced
2023-06-28
Critical
6
Units
810,580

Why these products were recalled

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–50 of 120

  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Medtronic recalls implantable cardioverter defibrillators due to rare potential for reduced or no-energy output during high-voltage therapy, which could prevent life-saving defibrillation.

    Product
    ICD-VR DVBB1D4 EVERA XT DF4 US, Model Number DVBB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Energy During Therapy

    Medtronic implantable cardioverter defibrillators (CRT-Ds) may fail to deliver adequate energy during high-voltage treatment due to a manufacturing defect in the feedthrough component, affecting 373 units.

    Product
    CRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail to Deliver Therapy

    Certain Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy due to a defective feedthrough component affecting 71 units.

    Product
    ICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy devices due to a rare potential for reduced or no energy output during high voltage therapy. Approximately 37,547 units are affected.

    Product
    ICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Therapy Failure

    Medtronic's ICD COBALT defibrillators may fail to deliver critical therapy due to a manufacturing defect in a power delivery component. Patients should contact their physician to determine if affected.

    Product
    ICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may experience reduced or no energy output during high-voltage therapy, potentially preventing treatment of life-threatening heart rhythms.

    Product
    ICD-DR DDMB2D1 EVERA MRI XT OUS DF1, Model Number DDMB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a specific feedthrough manufacturing defect. Approximately 8,219 units were affected worldwide.

    Product
    ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may fail to deliver therapy due to a defect in the electrical feedthrough. The devices risk reduced or no energy output during high-voltage treatment.

    Product
    ICD-VR VISIA MRI AF S US/OUS DF4, Model Number DVFC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential High-Voltage Therapy Failure

    Medtronic is recalling 664 implantable cardioverter defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy. The defect could prevent critical defibrillation therapy when needed.

    Product
    CRTD DTMC1D1 COMPIA MRI US DF1, Model Number DTMC1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy During Critical Events

    Certain Medtronic implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough manufacturing defect. 523 units distributed worldwide are affected.

    Product
    CRT-D DTBB2QQ VIVA QUAD S IS4/DF4 INTL, Model Number DTBB2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardiac Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardiac defibrillators with specific glassed feedthrough components may deliver reduced energy during high-voltage therapy. This could prevent the device from functioning when needed in cardiac emergencies.

    Product
    CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    FDA Recalls Medtronic Implantable Defibrillators Due to Potential Output Risk

    Medtronic's ICD-VR VISIA AF S implantable defibrillators may fail to deliver energy during high-voltage therapy due to a feedthrough component defect. The FDA Class I recall affects 194 units distributed nationwide and worldwide.

    Product
    ICD-VR VISIA AF S US/OUS DF4, Model Number DVAC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced or No Therapy Output

    Certain Medtronic implantable cardioverter defibrillators and cardiac resynchronization defibrillators may fail to deliver energy during high-voltage therapy due to a specific feedthrough design issue. The recall affects 3,801 units distributed in the US and worldwide.

    Product
    ICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-28

    Medtronic Implantable Cardiac Defibrillators Risk Reduced Energy During Therapy

    Medtronic implantable cardiac defibrillators may deliver reduced or no energy during defibrillation therapy due to a feedthrough defect. Approximately 11,239 devices distributed worldwide are affected.

    Product
    ICD-DR DDMD3D4 PRIMO MRI, Model Number DDMD3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Critical Therapy

    Medtronic implantable defibrillators may produce reduced or no energy output during high-voltage therapy delivery. Affected devices were manufactured with a specific feedthrough component.

    Product
    CRTD DTMC2QQ COMPIA MRI QUAD OUS DF4, Model Number DTMC2QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardiac Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic's ICD-VR DVME3D4 defibrillators are recalled due to a rare risk of reduced or no energy output during high-voltage therapy. The defect involves a specific feedthrough component and affects approximately 14,989 units distributed nationwide and worldwide.

    Product
    ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Medtronic implantable cardioverter defibrillators may fail to deliver critical therapy due to a feedthrough defect. The defect can cause reduced or absent energy output during high-voltage therapy in 8,602 affected units distributed worldwide.

    Product
    CRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Heart Defibrillators Risk Therapy Delivery Failure

    Medtronic CRT-D and ICD implantable defibrillators may fail to deliver therapy due to a rare defect in a specific feedthrough component. The recalled devices carry a potential for reduced or no-energy output during high-voltage heart rhythm treatment.

    Product
    CRT-D DTBA2D4 VIVA XT IS1/DF4 INTL, Model Number DTBA2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Deliver Reduced Energy During Therapy

    Medtronic ICDs and CRT-Ds with a specific feedthrough may deliver reduced or no energy during high voltage therapy. Approximately 2,669 units are affected in the US and worldwide.

    Product
    CRTD COBALT XT HF QUAD MRI IS4 DF1, Model Number DTPA2Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Risk Reduced or No Energy Output During Therapy

    Certain Medtronic implantable defibrillators may fail to deliver high voltage therapy due to a feedthrough defect. The FDA Class I recall affects 2,282 units distributed nationwide and worldwide.

    Product
    ICD-DR DDBB2D1 EVERA XT IS1/DF1 INTL, Model Number DDBB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Rare Therapy Delivery Failure

    Certain Medtronic implantable defibrillators may fail to deliver high voltage therapy in rare cases due to a manufacturing defect. The recall affects approximately 61,404 units distributed worldwide.

    Product
    ICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable defibrillators may fail to deliver high-voltage therapy due to defect

    Medtronic's EVERA XT implantable defibrillators may fail to deliver high-voltage therapy in certain models manufactured with a defective feedthrough. Patients should consult their cardiologist.

    Product
    ICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable defibrillators with a specific feedthrough defect may rarely experience reduced or no-energy output during high-voltage therapy. Patients should contact their healthcare provider for evaluation.

    Product
    ICD-VR DVMB1D4 EVERA MRI XT DF4 US, Model Number DVMB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states