FDA (Devices) recall action 92328

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalled 120 products on 2023-06-28

The FDA (Devices) publishes a recall like this one as 120 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 120 pages.

Products
120
Announced
2023-06-28
Critical
6
Units
810,580

Why these products were recalled

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 120

  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillator may lose energy output during therapy

    Medtronic recalls approximately 5,622 implantable cardioverter defibrillators that may produce reduced or no energy output during high-voltage therapy due to a specific feedthrough component defect.

    Product
    ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Certain Medtronic implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a specific feedthrough defect. The 4,767 affected units were distributed nationwide and worldwide.

    Product
    CRTD DTMB2Q1 AMPLIA MRI QUAD OUS DF1, Model Number DTMB2Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Reduced Therapy Output

    Medtronic is recalling 14 units of CRTD VIVA QUAD XT implantable cardioverter defibrillators due to rare potential for reduced or no energy output during high voltage therapy. Affected devices were manufactured with a specific feedthrough defect.

    Product
    CRTD VIVA QUAD XT CUST PARY GOLD, Model Number DTBA1QQPG; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators at Risk of Reduced Energy Output

    Medtronic ICDs and CRT-Ds with a specific feedthrough defect may produce reduced or no energy output during high-voltage therapy. The recall involves approximately 4,546 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB2D1 VISIA AF MRI XT OUS DF1, Model Number DVFB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic recalls certain ICD and CRT-D defibrillators due to a rare potential for reduced or no-energy output during high-voltage therapy. Affected devices were manufactured with a specific glassed feedthrough component.

    Product
    ICD-VR DVME3D1 MIRRO MRI, Model Number DVME3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-06-28

    Medtronic CRT-D Defibrillators: Rare Energy Output Failure Risk During Therapy

    Medtronic implantable cardioverter defibrillators (CRT-D) may deliver reduced or no energy during high-voltage therapy due to manufacturing defect in the feedthrough component. FDA Class I recall affects 182 units.

    Product
    CRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic CRT-D and ICD Defibrillators Recalled for Potential Therapy Output Failure

    Medtronic has recalled 891 implantable cardioverter defibrillators (CRT-D) worldwide due to a rare potential for reduced or absent energy output during defibrillation therapy. Units manufactured with a specific feedthrough component are affected.

    Product
    CRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Medtronic implantable defibrillators may fail to deliver therapy during critical cardiac events due to a feedthrough defect. Approximately 25,058 units are affected nationwide and worldwide.

    Product
    ICD-VR DVMC3D4 EVERA MRI S DF4 GLOB, Model Number DVMC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic implantable defibrillators may fail to deliver therapy

    Medtronic implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators with specific feedthrough components may fail to deliver adequate high-voltage therapy output.

    Product
    CRTD DTMC2D1 COMPIA MRI OUS DF1, Model Number DTMC2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling certain ICD and CRT-D devices that may fail to deliver high-voltage therapy due to a manufacturing defect in the device feedthrough. The recall affects 2,018 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic ICD defibrillators may fail to deliver critical therapy

    Medtronic recalled certain implantable defibrillators that may fail to deliver electrical therapy during critical heart rhythm events due to a manufacturing defect in the device feedthrough.

    Product
    ICD-VR VISIA AF XT OUS DF4, Model Number DDVAB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable defibrillators may deliver reduced or no energy during therapy

    Medtronic implantable defibrillators and cardiac resynchronization devices with a specific glassed feedthrough may fail to deliver full energy during therapy. Approximately 8,575 units affected worldwide.

    Product
    CRTD DTMB2D4 AMPLIA MRI OUS DF4, Model Number DTMB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Reduced Energy Output

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a manufacturing issue. The recall affects 1,354 units distributed in the United States and internationally.

    Product
    ICD-VR DVBB2D1 EVERA XT IS1/DF1 INTL, Model Number DVBB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-28

    Implantable Cardiac Defibrillators May Deliver Reduced Energy Output

    Certain Medtronic cardiac defibrillators may deliver reduced or no energy during high voltage therapy. No injuries or deaths have been reported.

    Product
    CRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail During High-Voltage Therapy

    Some Medtronic implantable defibrillators contain a specific feedthrough that may rarely cause reduced or no energy output during high-voltage therapy. The recall affects approximately 4,151 units distributed nationwide.

    Product
    CRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic's implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a feedthrough defect. Approximately 6,821 devices are affected worldwide.

    Product
    ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential High Voltage Therapy Failure

    Medtronic implantable cardioverter defibrillators (ICDs) with a specific feedthrough component may have reduced or absent energy output during high voltage therapy. The FDA Class I recall affects 12,099 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB1D1 VISIA AF MRI US DF1, Model Number DVFB1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Risk Loss of Therapy from Feedthrough Defect

    Medtronic recalls certain CRT-D VIVA QUAD XT GOLD implantable defibrillators due to a rare risk of reduced or no electrical output during therapy. Affected devices contain a specific glassed feedthrough component.

    Product
    CRT-D VIVA QUAD XT GOLD CTD, Model Number DTBA2QQG; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic implantable cardioverter defibrillators (ICDs) and CRT-Ds with specific feedthrough designs may produce reduced or insufficient energy during high voltage therapy, affecting the device's ability to treat heart arrhythmias.

    Product
    ICD-DR DDMB2D4 EVERA MRI XT IS-1/DF4 INT, Model Number DDMB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable defibrillators risk energy output failure during high voltage therapy

    Certain Medtronic cardiac defibrillators may produce reduced or no energy output during high voltage therapy. The defect involves a specific feedthrough component used in the manufacturing process.

    Product
    CRT-D DTBA2D1 VIVA XT IS1/DF1 INTL, Model Number DTBA2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators Recalled for Rare Risk of Energy Output Failure

    Medtronic implantable defibrillators may deliver reduced or no energy during therapy due to a rare electrical defect. Approximately 1,914 units with a specific feedthrough component are affected and distributed worldwide.

    Product
    CRTD COBALT XT HF MRI IS1 DF4, Model Number DTPA2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially compromising treatment effectiveness.

    Product
    ICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardiac Defibrillators Risk Therapy Delivery Failure

    Medtronic is recalling certain CRT-D and ICD models due to a potential feedthrough defect that may prevent energy delivery during high-voltage therapy. The FDA Class I recall affects 435 units distributed in the United States and worldwide.

    Product
    CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail During High-Voltage Cardiac Therapy

    Medtronic implantable defibrillators may fail to deliver energy during high-voltage cardiac therapy due to a defective glassed feedthrough component. The recall affects 2,968 units distributed nationwide and worldwide.

    Product
    CRTD COBALT HF MRI IS1 DF4, Model Number DTPB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states