Implantable Cardioverter Defibrillators Risk Energy Output Failure
Medtronic Inc. is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) due to a rare potential for reduced or no-energy output during high voltage therapy. Affected devices contain a specific glassed feedthrough.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: FDA Class I classification mandates a minimum Severe rating. Although no injuries or deaths are reported in the source, the potential for complete treatment failure in a critical implantable cardiac device justifies this severity level.
Plain-English summary
Medtronic Inc. is recalling certain implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds), model DTBA2QQ VIVA QUAD XT, that were manufactured with a specific glassed feedthrough. The recall affects 2,931 units distributed nationwide and worldwide.
The recalled devices carry a rare potential for reduced or no-energy output during high voltage (HV) therapy, typically in the 0-12J energy range. This malfunction could prevent the device from delivering appropriate shock therapy when medically necessary.
Patients with affected devices should contact their healthcare provider or cardiologist to verify whether their device is included in this recall. Medtronic has identified specific lot serial numbers to facilitate identification.
The recalled product
- Product
- CRT-D DTBA2QQ VIVA QUAD XT IS4/DF4 INTL, Model Number DTBA2QQ; Implantable Cardioverter Defibrillators
- Affected units
- 2,931
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- GTIN 00643169720169
- Lot Serial Numbers: BLG624311S
- BLG624755S
- BLG625177S
- BLG625181S
- BLG625182S
- BLG625188S
- BLG625193S
- BLG625324S
- BLG625325S
- BLG625434S
- BLG625596S
- BLG625598S
- BLG625606S
- BLG625608S
- BLG625636S
- BLG625647S
- BLG625702S
- BLG625740S
- BLG625741S
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 120 recalled by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 120 products in this recall →Related recalls
Same manufacturer · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
See all →- SevereImplantable Defibrillators Risk Therapy Failure Due to Feedthrough Defect
FDA (Devices) · 2023-06-28
- SevereMedtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure
FDA (Devices) · 2023-06-28
- SevereMedtronic Implantable Cardiac Defibrillators Risk Reduced Energy During Therapy
FDA (Devices) · 2023-06-28
- SevereMedtronic Implantable Cardioverter Defibrillator may lose energy output during therapy
FDA (Devices) · 2023-06-28
- SevereImplantable Cardioverter Defibrillators at Risk of Reduced Energy Output
FDA (Devices) · 2023-06-28
Same hazard · device malfunction
See all →- SevereJenaValve Trilogy Delivery Catheter Recalled for Sleeve Liner Separation
FDA (Devices) · 2026-10-21
- SevereJenaValve Trilogy Delivery Catheter Recalled for Sleeve Liner Separation
FDA (Devices) · 2026-10-21
- SevereJenaValve Trilogy Delivery Catheter Recalled for Sleeve Liner Separation
FDA (Devices) · 2026-10-21
- ModerateMedline recalls CONMED AirSeal Access Port kits due to pressure alert issues
FDA (Devices) · 2026-10-21
- ModerateMedline Recalls Convenience Kits Containing CONMED AirSeal Access Ports
FDA (Devices) · 2026-10-21
Similar recalls
- SevereCobalt Implantable Cardioverter Defibrillators recalled for reduced shock energy
FDA (Devices) · 2022-08-17
- SevereMedtronic Cobalt Implantable Defibrillators Recalled Due to Reduced Shock Energy Defect
FDA (Devices) · 2022-08-17
- SevereCobalt Implantable Defibrillators May Deliver Reduced Shock Energy
FDA (Devices) · 2022-08-17
- SevereMedtronic Recalls 78 Viva Quadripolar Implantable Cardioverter Defibrillators
FDA (Devices) · 2016-08-31
- SevereCobalt Implantable Cardioverter Defibrillators May Deliver Reduced Shock Energy
FDA (Devices) · 2022-08-17
- SevereCrome Implantable Cardioverter Defibrillators may deliver reduced shock energy
FDA (Devices) · 2022-08-17