FDA (Devices) recall action 92328

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalled 120 products on 2023-06-28

The FDA (Devices) publishes a recall like this one as 120 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 120 pages.

Products
120
Announced
2023-06-28
Critical
6
Units
810,580

Why these products were recalled

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

76–100 of 120

  • SevereFDA (Devices)··2023-06-28

    Implantable Heart Defibrillators May Fail to Deliver Therapy

    Medtronic implantable heart defibrillators may fail to deliver therapy during emergencies. The rare defect affects 510 devices distributed nationwide and worldwide.

    Product
    ICD CROME VR MRI IS1 DF1, Model Number DVPC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic CRT-D Implants Recalled for Rare Energy Output Failure Risk

    Medtronic is recalling certain CRT-D (cardiac resynchronization therapy defibrillator) models due to a rare risk of reduced or no energy output during high-voltage therapy. The FDA has classified this as a Class I recall affecting 1,473 units.

    Product
    CRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic CRT-D Defibrillators Recalled for Potential Therapy Output Failure

    Medtronic implantable cardiac defibrillators (CRT-D) may deliver reduced or no energy during critical high-voltage therapy due to a feedthrough defect. The recall affects 193 units that could fail to deliver lifesaving defibrillation shocks when needed.

    Product
    CRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic implantable defibrillators recalled for potential therapy delivery failure

    Medtronic's ICD and CRT-D devices may fail to deliver high-voltage therapy due to a defect in a specific feedthrough component. Patients should contact their healthcare provider immediately to discuss the risk.

    Product
    ICD-VR DVBC3D4 EVERA S DF4 GLOB, Model Number DVBC3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapy Due to Feedthrough Defect

    Medtronic's ICD-VR DVMB2D4 implantable defibrillators may fail to deliver electrical therapy due to a defect in the feedthrough component. The recall affects 6,477 units distributed worldwide.

    Product
    ICD-VR DVMB2D4 EVERA MRI XT DF4 INTL, Model Number DVMB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators May Fail To Deliver Therapy; Class I Recall

    Medtronic implantable cardioverter defibrillators (ICDs) may fail to deliver high-voltage therapy due to a manufacturing defect in a specific feedthrough component. 447 units are affected worldwide.

    Product
    ICD-VR DVBB2D4 EVERA XT DF4 INTL, Model Number DVBB2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic is recalling 522 implantable cardioverter defibrillators that may deliver reduced or no energy during high-voltage therapy due to a feedthrough defect.

    Product
    ICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardiac Devices Risk Therapy Failure From Feedthrough Defect

    Certain Medtronic implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a defective feedthrough component. The FDA classified this as a Class I recall affecting 2,884 devices.

    Product
    CRTD COBALT HF MRI IS1 DF1, Model Number DTPB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Potential Therapy Failure

    Medtronic is recalling 661 CRT-D implantable defibrillators due to a rare manufacturing defect that may reduce or eliminate energy output during therapy delivery. Affected devices may fail to deliver necessary electrical treatment during cardiac emergencies.

    Product
    CRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough design issue. This could prevent life-saving electrical therapy from being delivered.

    Product
    CRTD DTMC2D4 COMPIA MRI OUS DF4, Model Number DTMC2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Reduced Energy Output During Therapy

    Medtronic's implantable cardioverter defibrillators may produce reduced or no energy during high voltage therapy due to a feedthrough defect. A total of 1,076 units are affected worldwide.

    Product
    CRT-D DTBB2D1 VIVA S IS1/DF1 INTL, Model Number DTBB2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver High Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators may deliver reduced or no energy during high voltage therapy due to a defective feedthrough. This defect could prevent the device from properly treating life-threatening heart rhythms.

    Product
    ICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-28

    Implantable Cardiac Defibrillators at Risk of Therapy Delivery Failure

    Medtronic implantable defibrillators may fail to deliver therapy due to a feedthrough defect. 2,173 affected units were distributed nationwide and worldwide.

    Product
    CRTD DTMB1Q1 AMPLIA MRI QUAD US DF1, Model Number DTMB1Q1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    ICD COBALT XT Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic recalled 22,114 ICD COBALT XT defibrillators due to a rare potential for reduced or no energy output during high-voltage therapy delivery caused by a specific feedthrough defect.

    Product
    ICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Cardioverter Defibrillators Risk Failure to Deliver Therapy

    Medtronic implantable cardioverter defibrillators with a specific feedthrough defect may rarely fail to deliver high-voltage therapy. Approximately 34,350 devices distributed worldwide are affected.

    Product
    ICD-VR DVFB1D4 VISIA MRI AF US DF4, Model Number DVFB1D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Cardioverter Defibrillators Recalled: Potential Reduced or No-Energy Output

    Certain Medtronic implantable cardioverter defibrillators may fail to deliver proper energy output during high-voltage therapy due to a feedthrough defect. Patients with affected devices should contact Medtronic immediately.

    Product
    CRTD DTMC1QQ COMPIA MRI QUAD US DF4, Model Number DTMC1QQ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators With Defective Feedthrough May Fail During Critical Therapy

    Medtronic is recalling 14,270 implantable cardioverter defibrillators with a specific feedthrough that may produce reduced or no energy output during high-voltage therapy. The affected devices may fail to deliver critical treatment when needed.

    Product
    ICD-VR DVMD3D4 PRIMO MRI, Model Number DVMD3D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillator Models May Fail to Deliver Shock Therapy

    Medtronic implantable cardiac defibrillators may fail to deliver electrical therapy during treatment due to a defective feedthrough component. Approximately 1,878 devices are affected.

    Product
    CRTD COBALT HF QUAD MRI IS4 DF1, Model Number DTPB2Q1 ; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapeutic Shock

    Medtronic implantable cardioverter defibrillators with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially failing to deliver life-saving shocks.

    Product
    CRTD DTMA2D4 CLARIA MRI OUS DF4, Model Number DTMA2D4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling 289 implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators due to a rare potential defect that may prevent proper high-voltage therapy delivery.

    Product
    ICD-VR DVEX3E4 EV ICD EV4 PIVOTAL, Model Number DVEX3E4; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillators May Have Reduced Energy Output During Therapy

    Medtronic is recalling approximately 6,525 implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) that may produce reduced or no energy output during high-voltage therapy, potentially preventing treatment delivery.

    Product
    ICD-VR DVMC3D1 EVERA MRI S OUS/US DF1, Model Number DVMC3D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Implantable Defibrillator Recall: Risk of Failed High-Voltage Therapy

    Medtronic implantable defibrillators may fail to deliver therapy due to a defective feedthrough component. 207 units recalled nationwide and worldwide.

    Product
    CRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-28

    Medtronic Implantable Defibrillators May Fail to Deliver Therapy

    Medtronic is recalling certain implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators due to a rare risk of reduced or no energy output during high voltage therapy.

    Product
    CRTD DTBA2D1G VIVA XT IS1 DF1 OUS GOLD, Model Number DTBA2D1G; Implantable Cardioverter Defibrillators
    Category
    Distribution
    0 states