The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

35451–35475 of 40489

  • HighFDA (Devices)·Z-2062-2014·2014-07-23

    Instrumentation Laboratory Recalls ACL TOP CTS Coagulation Instruments

    Instrumentation Laboratory is recalling 572 ACL TOP CTS automated coagulation laboratory instruments due to a potential for sample misidentification.

    Product
    ACL TOP CTS, automated coagulation laboratory instrument.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2070-2014·2014-07-23

    Excelsior Medical Recalls Sterile Triple Lead Tubing Sets Due to Leak Risk

    Excelsior Medical is recalling sterile triple lead tubing sets because a molding defect may allow fluid leaks or air entry during use.

    Product
    Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2029-2014·2014-07-23

    Terumo Sarns Centrifugal Pump Recalled for Pumphead Leaks

    Terumo is recalling specific lots of Sarns Disposable Centrifugal Pumps because leaks were detected on the pumphead during set-up and priming.

    Product
    Sarns Disposable Centrifugal Pump with X-Coating, 164275X, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2064-2014·2014-07-23

    Instrumentation Laboratory Recalls ACL TOP 500 CTS Coagulation Instruments

    Instrumentation Laboratory is recalling ACL TOP 500 CTS automated coagulation laboratory instruments due to a potential for sample misidentification.

    Product
    ACL TOP 500 CTS, automated coagulation laboratory instrument.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2045-2014·2014-07-23

    Smith & Nephew Recalls SUTUREFIX Ultra Anchor XL Due to Sterility Breach

    Smith & Nephew is recalling 259 units of SUTUREFIX Ultra Anchor XL devices because a breach in the sterile barrier compromised the device's sterility.

    Product
    SUTUREFIX Ultra Anchor XL with one (#2) Ultrabraid Blue Product Code: 72203841 Fastener, fixation, nondegradable, soft tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2028-2014·2014-07-23

    Terumo Sarns Centrifugal Pump Recalled for Pumphead Leaks

    Terumo is recalling Sarns Disposable Centrifugal Pumps because leaks were detected on the pumphead during set-up and priming.

    Product
    Sarns Disposable Centrifugal Pump without X-Coating,164275, is a sterile, single use device. For use in cardiopulmonary bypass procedures.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2048-2014·2014-07-23

    Smith & Nephew Recalls SUTUREFIX Ultra S Suture Due to Sterility Breach

    Smith & Nephew is recalling SUTUREFIX Ultra S sutures because the sterile barrier is compromised, posing a risk of infection.

    Product
    SUTUREFIX Ultra S with 1 #2 Ultrabraid Suture (blue/white) Product Code: 72203853 Fastener, fixation, nondegradable, soft tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2055-2014·2014-07-23

    Stryker Communications Recalls SHAPE Arm Brackets for Boom Systems

    Stryker Communications is recalling 953 SHAPE Arm brackets used in operating rooms because the monitor mount may detach if specific hardware is missing.

    Product
    BAM (SHAPE) ARM, LEFT, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as component of kit part number 0682000976, and the Arm…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2053-2014·2014-07-23

    Stryker SHAPE Arm Dual SH2 Recalled for Potential Monitor Detachment

    Stryker Communications is recalling SHAPE Arm Dual SH2 brackets because the monitor mount may detach if specific hardware is missing.

    Product
    SHAPE ARM, DUAL SH2, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400254 and the Arm…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2056-2014·2014-07-23

    Stryker Communications Recalls SHAPE Arm Brackets for Potential Monitor Detachment

    Stryker Communications is recalling 953 SHAPE Arm brackets because the monitor mount may detach if a lock washer and screw are missing.

    Product
    BAM (SHAPE) ARM, RIGHT, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682000977, and the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2061-2014·2014-07-23

    Instrumentation Laboratory Recalls ACL TOP 700 Coagulation Instruments

    Instrumentation Laboratory is recalling 633 ACL TOP 700 automated coagulation laboratory instruments due to a potential for sample misidentification.

    Product
    ACL TOP 700, automated coagulation laboratory instrument.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2063-2014·2014-07-23

    Instrumentation Laboratory Recalls ACL TOP 700 LAS Coagulation Instruments

    Instrumentation Laboratory is recalling 99 ACL TOP 700 LAS automated coagulation laboratory instruments due to a potential for sample misidentification.

    Product
    ACL TOP 700 LAS, automated coagulation laboratory instrument.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2059-2014·2014-07-23

    Mindray V-Series Patient Monitoring System Recalled for ECG Interpretation Errors

    Mindray is recalling 113 V-Series Patient Monitoring Systems due to software issues causing false ECG measurements and incorrect interpretive statements.

    Product
    V-Series Patient Monitoring System, Manufactured for Mindray DS USA, Inc. Used for monitoring of human physiological parameters. The V-Series Monitor also has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-2049-2014·2014-07-23

    Smith & Nephew Recalls SUTUREFIX Ultra S Sutures Due to Sterility Breach

    Smith & Nephew is recalling 763 units of SUTUREFIX Ultra S Double Loaded Sutures because the sterile barrier was breached, compromising the device's sterility.

    Product
    SUTUREFIX Ultra S Double Loaded Suture Product Code: 72203854 Fastener, fixation, nondegradable, soft tissue
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2065-2014·2014-07-23

    Instrumentation Laboratory Recalls ACL TOP 700 CTS Coagulation Instruments

    Instrumentation Laboratory is recalling 266 ACL TOP 700 CTS automated coagulation laboratory instruments due to a potential for sample misidentification.

    Product
    ACL TOP 700 CTS, automated coagulation laboratory instrument.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2069-2014·2014-07-23

    Axcentria Recalls Calcium Hydroxide Powder Due to Missing 510(k) Approval

    Axcentria Pharmaceuticals is recalling one lot of Calcium Hydroxide Powder because it was distributed without FDA approval and lacks usage instructions.

    Product
    Calcium Hydroxide Powder, USP For use in root canal therapy by health care professional. Product Application or Usage: Provides bactericidal effect on carious dentine. Paste delivered to root apex provides pulp insulation. Protects the vital pulp from infections and further…
    Category
    Medical Device
    Distribution
    2 states
  • ModerateFDA (Devices)·Z-2043-2014·2014-07-23

    Siemens Symbia E Camera System Recalled for Loose Rear Casters

    Siemens Medical Solutions USA is recalling 531 Symbia E camera systems because the collimator cart rear casters may become loose.

    Product
    Symbia E camera system. Emission Computed Tomography System used to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588 keV. Nuclear…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2078-2014·2014-07-23

    Dental EZ Group Recalls 45 Mislabelled Titan T 5K Dental Motors

    Dental EZ Group is recalling 45 Titan T 5K dental motors because the housing was incorrectly laser marked as a 20K motor.

    Product
    Titan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety of dental procedures including caries removal, finishing, polishing, pin setting and prophy.The motor has a speed range of 100 to 5000 rpm capability.
    Category
    Medical Device
    Distribution
    7 states
  • ModerateFDA (Devices)·Z-2077-2014·2014-07-23

    Spacelabs Ultraview SL Multigas Module Recalled for Invalid MAC Display

    Spacelabs Healthcare is recalling 8 Ultraview SL Multigas Modules because the software displays an invalid Minimum Alveolar Concentration value.

    Product
    Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves…
    Category
    Medical Device
    Distribution
    1 state
  • ModerateFDA (Devices)·Z-2057-2014·2014-07-23

    Remel Inc. Recalls Thermo Scientific VersaTREK REDOX 1 In Vitro Diagnostic Product

    Remel Inc. is recalling specific lots of Thermo Scientific VersaTREK REDOX 1 in vitro diagnostic products because they failed to provide positive signals for Streptococcus pneumoniae.

    Product
    Thermo Scientific VersaTREK REDOX 1, Ref 7102-44, 80ml with Stir Bar (O2 aerobic), 50 bottles/box, IVD. The firm name on the label is Remel Inc., Lenexa, KS.
    Category
    Medical Device
    Distribution
    8 states
  • ModerateFDA (Devices)·Z-2073-2014·2014-07-23

    Laser Peripherals Recalls Mislabelled Bare Laser Fibers Due to Date Error

    Laser Peripherals LLC is recalling 25 units of DBLF-60-2 Bare Laser Fibers because a sterility expiration date was misprinted as '2016-16' instead of '2016-06'.

    Product
    LP,Surgical Fibers, Bare Laser Fiber 600 micron, DBLF-60-2, Sterile EO. Laser surgical instrument for use in general and plastic surgery and in dermatology
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2050-2014·2014-07-23

    Zimmer Dental Tapered Screw-Vent Implant Recalled for Potential Vial Crack

    Zimmer, Inc. is voluntarily recalling a specific lot of dental implants due to a potential crack in the outer vial cap.

    Product
    Zimmer Dental Tapered Screw-Vent Implant, Catalog TSVT6B13, Lot 62284006
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2067-2014·2014-07-23

    Innovasis Recalls Excella II Pedicle Screws Due to Mislabeling

    Innovasis is recalling 32 units of Excella II Standard Pedicle Screws because the size and catalog number are mislabeled.

    Product
    Excella II Standard Pedicle Screw, Catalog No. E2S47525 and E2S47535, 4.75 mm x 25 mm. For spinal fixation. Packaged in heat-sealed plastic bag. Also provided in reusable steam sterilization tray as implant set.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2075-2014·2014-07-23

    Siemens syngo RT Dosimetrist 2.7 software update addresses data storage issue

    Siemens Medical Solutions is updating syngo RT Dosimetrist 2.7 software to fix a safety issue where structure names were not stored correctly.

    Product
    Siemens syngo RT Dosimetrist 2.7 system Product Usage: The intended use of syngo Suite for Oncology Systems is as an accessory to the linear accelerator system to aid and support in the planning and delivery of x-ray radiation for the therapeutic treatment of cancer.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2072-2014·2014-07-23

    Integra Recalls Camino ICP Catheters for MRI Safety Labeling

    Integra LifeSciences is recalling Camino Intracranial Pressure Monitoring Kits to update labeling indicating they are unsafe in MR environments.

    Product
    Integra Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters, Cat. No's. 110-4B, 110-4G, 110-4H, 110-4L
    Category
    Medical Device
    Distribution