Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

33951–33975 of 40489

  • HighFDA (Devices)·Z-0777-2015·2014-12-24

    Stryker Laparoscopic Scissors Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic hook scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, HOOK SCISSORS, 29CM, P/N 0250282049 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0810-2015·2014-12-24

    Siemens Biograph mCT Scanners Recalled for Potential Display Freeze

    Siemens Medical Solutions is recalling eight Biograph mCT S(64)-4R scanners due to a risk of system display freeze during CT interventional procedures.

    Product
    Biograph mCT S(64)-4R, System Material Number 10248672. The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0809-2015·2014-12-24

    Siemens Biograph mCT X-4R Recalled for Potential Display Freeze During Procedures

    Siemens Medical Solutions is recalling eight Biograph mCT X-4R scanners due to a possibility of system display freeze during CT interventional procedures.

    Product
    Biograph mCT X-4R, System Material Number 10248670. The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0811-2015·2014-12-24

    Siemens Biograph mCT S(40)-4R Recalled for Display Freeze Risk

    Siemens Medical Solutions is recalling eight Biograph mCT S(40)-4R scanners due to a potential system display freeze during CT interventional procedures.

    Product
    Biograph mCT S(40)-4R, System Material Number 10248671. The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-0759-2015·2014-12-24

    Stryker Laparoscopic Scissors Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of 3mm Peritoneal Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, PERITONEAL SCISSORS, 20CM, P/N 0250282001 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0452-2015·2014-12-24

    GE Healthcare Recalls Discovery IGS 740 Imaging Systems Due to Loss of Function

    GE Healthcare is recalling Discovery IGS 740 systems because they may lose real-time imaging or motion capabilities due to electromagnetic sensitivity.

    Product
    Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
    Category
    Medical Device
    Distribution
    5 states
  • HighFDA (Devices)·Z-0821-2015·2014-12-24

    Mitsubishi VERO Linear Accelerator Software Recall for Image Display Error

    Mitsubishi Heavy Industries is recalling four VERO Linear Accelerator systems in the US due to a software anomaly that may display incorrect image angles for patient positioning.

    Product
    VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.
    Category
    Medical Device
    Distribution
    4 states
  • ModerateFDA (Devices)·Z-0765-2015·2014-12-24

    Stryker Endoscopy Recalls 180,573 Cobra Grasper Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 Cobra Grasper instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, COBRA GRASPER, 20CM, P/N 0250282007 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0661-2015·2014-12-24

    Stryker Endoscopy Recalls Schertel Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of Schertel Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, SCHERTEL FORCEPS, BLUNT FENESTRATED, P/N 0250080072. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0676-2015·2014-12-24

    Stryker Debaney Forceps Recalled Due to Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 units of Debaney Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, DEBANEY FORCEPS, P/N 0250080314. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0626-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, RING HANDLE, P/N 0250080236. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0610-2015·2014-12-24

    Stryker Endoscopy Recalls Hasson Cannula Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Hasson Cannula units because the instructions for use do not support specific sterilization methods.

    Product
    PKG, HASSON CANNULA, 11MM AUTOMATIC VALVE, STOP COCK, P/N 0250080179. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0718-2015·2014-12-24

    Stryker Laparoscopic Manual Instruments Recalled for Sterilization Parameter Errors

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM ROTATING HANDLE, INSULATED SHAFT/ HANDLE 45CM LENGTH, P/N 0250080595 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0657-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Grasper Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic grasper forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, NARROW JAW, P/N 0250080287. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0685-2015·2014-12-24

    Stryker Laparoscopic Grasper Forceps Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of laparoscopic grasper forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, GRASPER FORCEPS, DELICATE, FENESTRATED, P/N 0250080322. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0785-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 3MM INSERT, MARYLAND, FORCEPS CURVED, 29CM, P/N 0250282058 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0751-2015·2014-12-24

    Stryker Laparoscopic Dissector Recall Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 laparoscopic dissectors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 45CM INSERT, RIGHT ANGLE DISSECTOR, LONG JAW, P/N 0250080761 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0617-2015·2014-12-24

    Stryker Endoscopy Recalls Bullet Forceps Due to Sterilization Documentation Issues

    Stryker Endoscopy is recalling 180,573 units of Bullet Forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BULLET FORCEPS, TAPERED, SERRATED, SPOON, P/N 0250080041. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0702-2015·2014-12-24

    Stryker Endoscopy Recalls Mini Metzenbaum Scissors Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of Mini Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, MINI METZENBAUM SCISSORS, CURVED, 45CM, P/N 0250080574 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0710-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Grasper Forceps Due to Sterilization Labeling Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic grasper forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, ATRAUMATIC GRASPER FORCEPS, DOUBLE ACTION 45CM, P/N 0250080585 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0706-2015·2014-12-24

    Stryker Laparoscopic Forceps Recalled for Sterilization Instructions Error

    Stryker Endoscopy is recalling 180,573 laparoscopic forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM INSERT, DISSECTING & GRASPER FORCEPS, SERRATED, 45CM, P/N 0250080579 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0593-2015·2014-12-24

    Stryker Endoscopy Recalls Bipolar Forceps Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 units of bipolar forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080120 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0643-2015·2014-12-24

    Stryker Metzenbaum Scissors Recalled Due to Sterilization Instructions

    Stryker Endoscopy is recalling 180,573 Metzenbaum Scissors because the instructions for use do not support specific sterilization methods.

    Product
    PKG, METZENBAUM SCISSORS, STRAIGHT, P/N 0250080264. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0750-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Dissecting Forceps Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic dissecting forceps because the instructions for use do not support specific sterilization methods.

    Product
    PKG, 5MM X 33CM INSERT, DISSECTING FORCEPS, P/N 0250080760 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF)…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-0659-2015·2014-12-24

    Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error

    Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.

    Product
    PKG, BIOPSY FORCEPS, PUNCH, P/N 0250080290. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
    Category
    Medical Device
    Distribution
    Distributed nationwide