Stryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Documentation Error
Stryker Endoscopy is recalling 180,573 units of laparoscopic manual instruments because the instructions for use do not support specific sterilization methods.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II and involves a documentation error regarding sterilization parameters, which fits the rubric for moderate severity involving minor labeling or documentation issues.
Plain-English summary
Stryker Endoscopy is recalling 180,573 units of PKG, BIOPSY FORCEPS, PUNCH, P/N 0250080290. These laparoscopic manual instruments are intended for cutting, holding, grasping, and manipulating tissue, organs, and surgical auxiliaries. The devices are distributed worldwide, including nationwide in the United States and in numerous international locations.
The recall is due to a documentation issue in the Instructions for Use (IFU 1000-401-070 Revision G or prior). The parameters provided in these instructions do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods. This is a Class II recall, indicating that use of the product could cause temporary or medically reversible health consequences.
Healthcare facilities and users should stop using the affected instruments and contact Stryker Endoscopy for further instructions on how to proceed with the recall.
The recalled product
- Product
- PKG, BIOPSY FORCEPS, PUNCH, P/N 0250080290. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which…
- Manufacturer
- Stryker Endoscopy
- Affected units
- 180,573
Distribution
Part of a larger recall action
This product is one of 207 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 207 products in this recall →Related recalls
Same manufacturer · Stryker Endoscopy
See all →- ModerateStryker PneumoSure Insufflators Recalled for Flow Rate Drop
FDA (Devices) · 2016-05-04
- SevereStryker Recalls SERFAS 90 Degree Energy Probes Due to Fragmentation
FDA (Devices) · 2015-07-01
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Labeling
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Laparoscopic Manual Instruments Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
- ModerateStryker Endoscopy Recalls Allis Forceps Due to Sterilization Instructions
FDA (Devices) · 2014-12-24
Same hazard · documentation error
See all →- LowMindray cPM Series Patient Monitor Manuals Recalled for Alarm Clarification
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- LowMindray Recalls Passport Series Patient Monitor Operator's Manuals
FDA (Devices) · 2026-10-14
- ModerateConMed Suture Anchor Recalled for Incorrect Patient Implant Card
FDA (Devices) · 2026-09-02
- ModerateHenry Schein Recalls Structur 2 SC Temporary Resin Cartridges for Incorrect Packaging
FDA (Devices) · 2026-09-02
Similar recalls
- ModerateStryker 12-Instrument Sterilization Tray Recalled for Validation Testing Gaps
FDA (Devices) · 2014-05-14
- ModerateStryker Arthroscope and Hardware Set Tray Recall for Sterilization Labeling
FDA (Devices) · 2012-07-18
- ModerateStryker Irrigator Reusable Tips Recalled for Incorrect Sterilization Instructions
FDA (Devices) · 2013-05-15
- SevereStryker Recalls ShapeMatch Cutting Guides Due to Internal Process Issues
FDA (Devices) · 2013-04-24
- ModerateStryker Irrigator Tips Recalled for Incorrect Sterilization Temperature Instructions
FDA (Devices) · 2013-05-15
- ModerateInstrumed Reusable Laparoscopic Instruments Recalled for Updated Safety Instructions
FDA (Devices) · 2014-10-08