The Recall Desk

Category

Medical device recalls

FDA recalls of medical devices including diagnostics, implants, and consumer health hardware.

About medical device recalls

Medical device recalls are the FDA's highest-volume stream. They cover implantable devices like pacemakers and joint replacements, hospital diagnostics and lab analyzers, surgical instruments, software-as-a-medical-device, and consumer-facing hardware like CPAP machines, blood-pressure cuffs, and glucose monitors. The volume comes from the wide range of products in scope and from the FDA's broad definition of 'corrective action' — software updates, IFU revisions, and labeling corrections all show up as recalls.

For Class I device recalls, the device's defect creates a reasonable risk of serious injury or death — these are the urgent ones. Class II covers issues that may cause temporary or medically reversible harm, often around accuracy or calibration. Class III is for products that don't comply with FDA rules but pose no real safety risk; in practice many software-update letters are Class III.

If you depend on a recalled device for ongoing care, don't disconnect or stop using it before talking to the prescribing clinician — interrupting therapy for a CPAP or a continuous glucose monitor can be more dangerous than the recall it's responding to. Almost every device recall offers a remedy that doesn't require you to surrender the device first.

32451–32475 of 40489

  • HighFDA (Devices)·Z-2057-2015·2015-08-05

    Hamilton-G5 Ventilators Recalled for Display Freeze Issue

    Hamilton Medical is recalling 248 Hamilton-G5 ventilators because the display can freeze, preventing users from operating the device while ventilation continues.

    Product
    Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2268-2015·2015-08-05

    OxiMax N-560 Pulse Oximeters Recalled for Display Segment Defects

    The FDA is recalling OxiMax N-560 Pulse Oximeters because missing display segments may cause users to misinterpret oxygen saturation and pulse rate data.

    Product
    OxiMax N-560 Pulse Oximeter. For continuous or spot check monitoring of functional arterial oxygen saturation (SpO2) and pulse rate.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2267-2015·2015-08-05

    OxiMax N-65 Pulse Oximeters Recalled for Display Segment Defects

    FDA is recalling OxiMax N-65 handheld pulse oximeters because missing display segments may cause users to misinterpret oxygen saturation and pulse rate data.

    Product
    OxiMax N-65 Handheld Pulse Oximeter. N65, N65-1, N65P, and N65P-1. For continuous or spot check monitoring of functional arterial oxygen saturation (SpO2) and pulse rate.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2270-2015·2015-08-05

    Prism Medical P-600 Portable Ceiling Lift Recalled for Sling Attachment Risk

    Prism Medical is recalling 409 P-600 Portable Ceiling Lifts because sling loops may not stay attached to the carry bar hooks.

    Product
    Prism Medical P-600 Portable Ceiling Lift; PN 303090. Patient lift with weight specification of 600 lbs (273 kg). Product Usage: The P-600 lift is a lifting aid used by health care professionals and those providing care in the home to lift, position, and transfer clients…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2269-2015·2015-08-05

    Prism Medical P-440 Portable Ceiling Lift Recalled for Sling Attachment Risk

    Prism Medical is recalling P-440 Portable Ceiling Lifts because sling loops may not stay attached to the carry bar hooks.

    Product
    Prism Medical P-440 Portable Ceiling Lift; PN 303070 and 303070R (refurbished); Patient lift with weight specification of 440 lbs (200 kg). Product Usage: The P-440 lift is a lifting aid used by health care professionals and those providing care in the home to lift…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2262-2015·2015-08-05

    Encore Medical Recalls Foundation Knee System Cut Blocks Due to Marking Error

    Encore Medical is recalling 10 units of Foundation Knee System cut blocks because markings are mis-oriented, which could lead to improper surgical cuts.

    Product
    FOUNDATION KNEE SYSTEM NON STERILE Femoral A/P Cut Block - Type II REF 800-01-139
    Category
    Medical Device
    Distribution
    2 states
  • HighFDA (Devices)·Z-2295-2015·2015-08-05

    Anspach 5MM Fluted Ball Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 79 Anspach 5MM Fluted Ball dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 5MM Fluted Ball, 2 mm Shaft Exposure; REF M-5B-2-G1. Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2314-2015·2015-08-05

    Siemens ACUSON S Family Ultrasound Systems Recalled for Measurement Error

    Siemens is recalling ACUSON S 1000, S 2000, and S 3000 ultrasound systems due to a potential measurement error in the Doppler manual trace tool that could lead to misdiagnosis.

    Product
    ACUSON S 1000, ACUSON S 2000, or ACUSON S 3000 ultrasound systems with software version C3, C3, C3, or C1. Model numbers: 10041461, 10440017  S 2000 system 10441730  S 3000 system 10441701  S 1000 system Radiology: The ultrasound imaging systems are intended for the…
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2257-2015·2015-08-05

    Philips CT Systems Recalled for Software and Hardware Defects

    Philips is recalling 82 Ingenuity Core CT systems due to software errors affecting image quality, dose control, and heart rate monitoring.

    Product
    Computed Tomography X-ray Systems Ingenuity Core. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2316-2015·2015-08-05

    Symmetry Surgical Disposable Towel Clamp Recalled for Sterility Assurance

    Symmetry Surgical is recalling 2,708 units of disposable towel clamps due to a lack of sterility assurance. The affected sterile products were distributed in the US, Germany, and Mexico.

    Product
    SYMMETRY SURGICAL DISPOSABLE TOWEL CLAMP, REF 19-4137, STERILE EO, Rx only. Apply clamps parallel to the skin with the angled jaws forming a "V" shape between clamp and drape.
    Category
    Medical Device
    Distribution
    18 states
  • HighFDA (Devices)·Z-2260-2015·2015-08-05

    Medtronic MiniMed Recalls 620G and 640G Insulin Pumps for Bolus Screen Issue

    Medtronic MiniMed is recalling 620G and 640G insulin pumps because the bolus screen may fail to timeout, displaying outdated dosing information.

    Product
    MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-2256-2015·2015-08-05

    Philips Brilliance iCT SP CT Systems Recalled for Software Defects

    Philips Medical Systems is recalling two Brilliance iCT SP CT systems due to software issues affecting image quality, dose control, and heart rate measurement.

    Product
    Computed Tomography X-ray Systems Brilliance iCT SP. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2255-2015·2015-08-05

    Philips Brilliance iCT CT Systems Recalled for Software and Hardware Defects

    Philips is recalling 38 Brilliance iCT CT systems due to software errors affecting dose control, heart rate measurement, and image artifacts.

    Product
    Computed Tomography X-ray systems Brilliance iCT, Philips Healthcare System. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2308-2015·2015-08-05

    Anspach 4mm Diamond Ball Dissection Tool Recall for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 246 units of the 4mm Diamond Ball dissection tool because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 4MM Diamond Ball, Extends 27MM; REF S-4D-L-G1 all lot numbers Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2258-2015·2015-08-05

    Philips CT Systems Recalled for Software and Hardware Defects

    Philips is recalling 139 Ingenuity Core 128 CT scanners due to software errors affecting image quality, dose control, and heart rate monitoring.

    Product
    Computed Tomography X-ray Systems Ingenuity Core 128. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-2315-2015·2015-08-05

    Tornier Aequalis Humeral Nail Targeting Jig Recalled for Mast Separation

    Tornier, Inc. is recalling 99 Aequalis Humeral Nail Targeting Jigs because the mast can separate from the boom, potentially affecting surgical alignment and screw fixation.

    Product
    Aequalis Humeral Nail Targeting Jig, Product Code: 9020060. Orthopedic manual surgical instrument.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2307-2015·2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 99 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual usage, potentially resulting in greater exposure than specified.

    Product
    ANSPACH -4MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-4DC-22-G1 Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2281-2015·2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 138 fluted drum dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Fluted Drum, Side Cut only 5MM x 14MM , REF L-5DRC-G1. Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2283-2015·2015-08-05

    Anspach Fluted Drum Dissection Tools Recalled for Inconsistent Shaft Exposure

    The Anspach Effort, Inc. is recalling 177 fluted drum dissection tools because the shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - Fluted Drum, Side Cut only, 6MM x 14MM REF L-6DRC-G1 Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2289-2015·2015-08-05

    Anspach Diamond Ball Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 72 Anspach Extra Coarse Diamond Ball tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure.

    Product
    ANSPACH - Extra Coarse Diamond Ball; 8MM REF L-8DX-G1 Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • HighFDA (Devices)·Z-2253-2015·2015-08-05

    OxySure Model 615 Oxygen Cartridges Recalled for Flow Rate Defect

    Oxysure Systems is recalling 1,005 Model 615 oxygen cartridges because they do not meet FDA flow rate standards.

    Product
    OxySure Model 615 disposable replacement cartridges, for portable oxygen generator.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • ModerateFDA (Devices)·Z-2301-2015·2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 95 G1 dissection tools because the specified shaft exposure may be greater than labeled when used with certain attachments.

    Product
    ANSPACH - 8MM x 12.5MM Fluted Barrel; REF M-8BA-G1 Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • ModerateFDA (Devices)·Z-2211-2015·2015-08-05

    EXP Pharmaceutical Services Recalls Unregistered Medical Device Exchange Surgical Gowns

    EXP Pharmaceutical Services Corp is recalling 10 units of Medical Device Exchange Surgical Gowns because the company failed to register as a medical device establishment or establish a quality system.

    Product
    Medical Device Exchange Surgical Gown - Flyte Hood Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number 408-800 EXP (repackager) Stryker (OEM) Person protection during surgery.
    Category
    Medical Device
    Distribution
    6 states
  • ModerateFDA (Devices)·Z-2288-2015·2015-08-05

    Anspach G1 Dissection Tools Recalled for Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 91 G1 dissection tools because the labeled shaft exposure may be inconsistent with actual use, potentially resulting in greater exposure than specified.

    Product
    ANSPACH - 8MM x 14MM Fluted Drum, Side Cut Only; REF L-8DRC-G1. Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states
  • ModerateFDA (Devices)·Z-2292-2015·2015-08-05

    Anspach G1 Dissection Tools Recalled Due to Inconsistent Shaft Exposure Labeling

    The Anspach Effort, Inc. is recalling 142 G1 dissection tools because the labeled shaft exposure may not be consistent with all listed attachments, potentially resulting in greater exposure.

    Product
    ANSPACH - 3MM Course Diamond Ball, 2 MM Shaft Exposure; REF M-3DC-2-G1 Cutting shaping bone including spine and cranium.
    Category
    Medical Device
    Distribution
    17 states