The Recall Desk

State

Vermont product recalls

72,940 recalls have nationwide distribution and so reach Vermont. 1,983 additional recalls listed Vermont specifically in their distribution scope.

What the numbers show

Total
1,983
Critical
199
Severe
409
Most recent
2026-09-02

Of the 1,983 recalls naming Vermont we have scored against our rubric, 199 (10%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 409 (21%) are Severe.

These figures cover only the 1,983 recalls whose distribution notice names Vermont specifically. The 72,940 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

4326–4350 of 74923

  • HighFDA (Devices)·Z-0498-2026·2025-11-19

    Long nasal feeding tubes recalled from one lot for possible occlusion

    Applied Medical Technology is recalling NutraGlide 6 F-8 F x 90 nasal feeding tubes with stylet and ENfit from one lot because the tube may be occluded.

    Product
    Applied Medical Technology, Inc., NutraGlide, 6 F- 8 F x 90 Nasal Feeding Tube with Stylet and ENfit, REF NFT-06090-I
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0471-2026·2025-11-19

    Five millimetre blunt tip sealers recalled over trapped residual material

    Medline is recalling two reprocessed LigaSure Blunt Tip sealer/dividers, 5 mm x 37 cm, because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal LigaSure Blunt Tip Sealer/Divider, Compatible with FT10 Generator, 5 mm x 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0486-2026·2025-11-19

    VATS lobectomy stapler boxes recalled over a malformed cover tube press

    Covidien is recalling two units of a VATS lobectomy box containing the Signia Small Diameter Curved Tip Intelligent Reload because internal components may move during firing.

    Product
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFN: BOX06170V4;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighNHTSA·25V800000·2025-11-19

    McNeilus Recalls 2024-2025 Volterra ZSL Trucks Due to Software Error

    McNeilus is recalling 2024-2025 Volterra ZSL side loader garbage trucks because a software error may cause grabber fingers to move without operator input, posing a risk of injury or crash.

    Product
    MCNEILUS — MCNEILUS VOLTERRA ZSL
    Category
    Vehicle
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0472-2026·2025-11-19

    Short blunt tip sealers recalled over residual material that may dislodge

    Medline is recalling 65 reprocessed LigaSure Blunt Tip sealer/dividers, 23 cm, because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline RenewalLigasure Blunt Tip Sealer/Divider, NanoCoated Compatible with FT10 Generator 23 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0452-2026·2025-11-19

    Lithotripsy transducers recalled over startup failures and handpiece temperature rise

    Olympus is recalling 1,528 ShockPulse Lithotripsy Transducers because the device may fail to start or stop shortly after starting, and the handpiece body may heat up during clinical use.

    Product
    Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse tra
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0453-2026·2025-11-19

    Patient monitors recalled over batteries that can lose backup power

    GE Healthcare is recalling 3,271 Carescape B450 patient monitors because certain batteries can lose the ability to hold charge, causing a loss of patient monitoring.

    Product
    Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE M
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0473-2026·2025-11-19

    Blunt tip sealers recalled over residual material trapped inside the device

    Medline is recalling 60,917 reprocessed LigaSure Blunt Tip sealer/dividers, 37 cm, because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal Ligasure Blunt Tip Sealer/Divider NanoCoated, Compatible with FT10 Generator 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0478-2026·2025-11-19

    Spinal fixation screws recalled for not meeting performance standards

    Orthofix is recalling 624 Virata preassembled polyaxial screws used in the Virata Spinal Fixation System because the screws do not meet performance standards.

    Product
    Virata Preassembled Polyaxial Screw, utilized in the Virata Spinal Fixation System. The following sizes are affected: 1. POLYAXIAL SCREW, CC, 5.5mm X 30mm; Item Number: TL1-10755030. 2. POLYAXIAL SCREW, CC, 5.5mm X 35mm; Item Number: TL1-10755035. 3. POLYAXIAL SCREW, CC
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0493-2026·2025-11-19

    Food allergen panels recalled over barcodes causing wrong tube scanning order

    Siemens Healthcare Diagnostics is recalling 28 units of Food Panel 3 Specific Allergen because barcode orientation causes the instrument to misread tube positions in the holder wedge.

    Product
    Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0485-2026·2025-11-19

    Surgical staplers recalled over a malformed cover tube press

    Covidien is recalling 3,263 Signia Small Diameter Curved Tip Intelligent Reloads because a manufacturing issue can allow internal components to move during firing, risking uncontrolled motion.

    Product
    Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0497-2026·2025-11-19

    Larger bore nasal feeding tubes recalled for possible occlusion

    Applied Medical Technology is recalling NutraGlide 6 F-8 F x 55 nasal feeding tubes with stylet and ENfit from one lot because the tube may be occluded.

    Product
    Applied Medical Technology, Inc., NutraGlide, 6 F- 8 F x 55 Nasal Feeding Tube with Stylet and ENfit, REF NFT-06055-I
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0454-2026·2025-11-19

    CARESCAPE ONE monitors recalled over batteries that fail to hold charge

    GE Healthcare is recalling 5,935 CARESCAPE ONE patient monitors because certain batteries can lose the ability to hold charge, causing a loss of patient monitoring.

    Product
    CARESCAPE ONE, Model no. 2087075-300 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Updated 3/19 to add CARESCAPE ONE MBZ101 2087075-001
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0182-2026·2025-11-19

    Dietary supplement recalled because it contains magnesium salicylate

    Professional Botanicals is recalling 430 bottles of Total Comfort capsules because the product contains magnesium salicylate.

    Product
    Product Name: Total Comfort. Package: 175cc HDPE Bottle. Type: Vegetable Capsule Count: 60, Dose: 2 Capsules. Product Sku: PB173.
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Food)·H-0183-2026·2025-11-19

    Tense-X supplement recalled because it contains magnesium salicylate

    Professional Botanicals is recalling 253 bottles of Tense-X capsules because the product contains magnesium salicylate.

    Product
    Product Name: Tense-X. Package: 175cc HDPE Bottle. Type: Vegetable Capsule. Count: 60. Dose: 2 Capsules. Product Sku: PB207
    Category
    Food
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0455-2026·2025-11-19

    Bedside patient monitors recalled over batteries that fail as backup power

    GE Healthcare is recalling 7,408 B105P, B125P, B105M, B125M and B155M patient monitors because certain batteries can lose the ability to hold charge, causing a loss of monitoring.

    Product
    B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 6160000-101 6160000-102 6160000-103 6160000-104 6160000-105 6160001-063 6160001-064 6160001-065 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms 3
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0467-2026·2025-11-19

    LigaSure Exact dissectors recalled over residual material that may dislodge

    Medline is recalling 4,226 reprocessed LigaSure Exact dissectors, 21 cm, because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal LigaSure Exact, Dissector without Nano-Coating 21cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical Cutting And Coagulation Device
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0470-2026·2025-11-19

    Small jaw sealers recalled over residual material that may dislodge

    Medline is recalling 18,117 reprocessed LigaSure Small Jaw instruments because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal LigaSure Small Jaw Instrument (Purple/White), Compatible with FT10 Generator 18.8 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0468-2026·2025-11-19

    Curved large jaw sealers recalled over residual material that may dislodge

    Medline is recalling one reprocessed LigaSure Impact Curved Large Jaw sealer/divider because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal LigaSure Impact Curved Large Jaw Sealer/Divider, Compatible with FT10 Generator, FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0476-2026·2025-11-19

    Reprocessed vessel sealers recalled over residual material that may dislodge

    Medline is recalling 57,244 reprocessed LigaSure Maryland Jaw sealer/dividers because certain areas of the device can trap residual material that may, in rare cases, become dislodged.

    Product
    Medline Renewal LigaSure Maryland Jaw Sealer/Divider NanoCoated, Compatible w/FT10 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0481-2026·2025-11-19

    Oxygen concentrator batteries recalled over a chip that can stop oxygen

    Philips Respironics is recalling 56 SimplyGo Mini standard battery kits because a defective chip can make the battery fail, shutting down the concentrator and stopping oxygen delivery.

    Product
    SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0483-2026·2025-11-19

    Fifteen millimetre snares recalled over holes compromising package sterility

    Olympus is recalling 18,174 Single Use Electrosurgical Snares SD-400U-15 because sterility may be compromised by small holes formed during packaging or transportation.

    Product
    Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use Electrosurgical Snare Model/Catalog Number: SD-U400U-15 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0490-2026·2025-11-19

    Reprocessed ultrasonic shears recalled because they may stick in the on position

    Stryker Sustainability Solutions is recalling 22,476 reprocessed HARMONIC FOCUS shears because a manufacturing issue may leave them stuck in the on position, causing continuous activation.

    Product
    HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0474-2026·2025-11-19

    Blunt tip sealers 44 cm recalled over trapped residual material

    Medline is recalling 1,446 reprocessed LigaSure Blunt Tip sealer/dividers, 44 cm, because certain areas of the device can trap residual material that may become dislodged.

    Product
    Medline Renewal Ligasure Blunt Tip Sealer/Divider NanoCoated, Compatible with FT10 Generator 44cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0482-2026·2025-11-19

    Electrosurgical snares recalled over small holes compromising package sterility

    Olympus is recalling 48,208 Single Use Electrosurgical Snares SD-400 because sterility may be compromised by small holes formed during packaging or transportation.

    Product
    Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare Model/Catalog Number: SD-400U-10 Product Description: These instruments have been designed to be used with an Olympus endoscope for the removal and/or cauterization of di
    Category
    Medical Device
    Distribution