The Recall Desk
HighFDA (Devices)·Z-0476-2026·Announced 2025-11-19

Reprocessed vessel sealers recalled over residual material that may dislodge

Medline is recalling 57,244 reprocessed LigaSure Maryland Jaw sealer/dividers because certain areas of the device can trap residual material that may, in rare cases, become dislodged.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: residual material that can dislodge from a surgical instrument is a foreign body hazard, and no injuries are reported.

Plain-English summary

The Medline Renewal LigaSure Maryland Jaw Sealer/Divider NanoCoated, compatible with FT10 37 cm, FT10 SW version 4.0.1.15 or lower, a reprocessed laparoscopic and endoscopic electrosurgical cutting and coagulation accessory, is being recalled.

Medline Renewal determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.

The recall covers 57,244 units across all lots, carrying UDI/DI 10888277719415 for each unit and 20888277719412 for the case. The product was distributed nationwide in the United States. The FDA has classified the action as Class II.

Material dislodging from an instrument during surgery could enter the patient. The enforcement record reports no injuries. Facilities should follow Medline's instructions.

The recalled product

Product
Medline Renewal LigaSure Maryland Jaw Sealer/Divider NanoCoated, Compatible w/FT10 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
Manufacturer
Medline Industries, LP
Hazard
  • residual-material
  • reprocessed-device
  • foreign-body

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • UDI/DI 10888277719415 (each)
  • 20888277719412 (case)
  • All Lots

Distribution

Distributed nationwide across the United States.