The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-0471-2026·Announced 2025-11-19

Five millimetre blunt tip sealers recalled over trapped residual material

Medline is recalling two reprocessed LigaSure Blunt Tip sealer/dividers, 5 mm x 37 cm, because certain areas of the device can trap residual material that may become dislodged.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Scored 3 under the risk-of-harm criterion: residual material that can dislodge from a surgical instrument is a foreign body hazard, and no injuries are reported.

Plain-English summary

The Medline Renewal LigaSure Blunt Tip Sealer/Divider, compatible with the FT10 generator, 5 mm x 37 cm, FT10 SW version 4.0.1.15 or lower, a reprocessed laparoscopic and endoscopic electrosurgical accessory, is being recalled.

Medline Renewal determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.

The recall covers 2 units across all lots, carrying UDI/DI 10888277680340 for each unit and 20888277680347 for the case. The product was distributed nationwide in the United States. The FDA has classified the action as Class II.

Material dislodging from an instrument during surgery could enter the patient. The enforcement record reports no injuries. Facilities should follow Medline's instructions.

The recalled product

Product
Medline Renewal LigaSure Blunt Tip Sealer/Divider, Compatible with FT10 Generator, 5 mm x 37 cm FT10 SW Version 4.0.1.15 or lower; Electrosurgical, Cutting & Coagulation Accessories, Laparoscopic & Endoscopic, Reprocessed
Affected units
2

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • UDI/DI 10888277680340 (each)
  • 20888277680347 (case)
  • All Lots

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 11 recalled by Medline Industries, LP under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 11 products in this recall →