The Recall Desk

State

Nebraska product recalls

73,716 recalls have nationwide distribution and so reach Nebraska. 4,580 additional recalls listed Nebraska specifically in their distribution scope.

What the numbers show

Total
4,580
Critical
713
Severe
1,345
Most recent
2026-10-28

Of the 4,580 recalls naming Nebraska we have scored against our rubric, 713 (16%) are Critical — deaths or serious injuries reported, or the agency’s own highest classification — and 1,345 (29%) are Severe.

These figures cover only the 4,580 recalls whose distribution notice names Nebraska specifically. The 73,716 nationwide recalls listed below reach every state, so counting them here would make every state page show the same numbers.

28726–28750 of 78296

  • HighFDA (Devices)·Z-0348-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 57 mm Number: 00-8065-554-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0331-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, Neutral, O.D. Cup with Spacers, 49 mm Item Number: 00-8065-546-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0390-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0341-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cup components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 55 mm Item Number: 00-8065-552-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0387-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Prosthesis Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 acetabular hip prosthesis cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0367-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Hip Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 acetabular hip cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 51 mm Number: 00-8065-648-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0332-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32mm, Neutral, O.D. Cup with Spacers, 49 mm Item Number: 00-8065-546-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0383-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0385-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Acetabular Cups Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 hip prosthesis acetabular cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0422-2021·2020-11-11

    3M Red Dot Monitoring Electrode Recall Due to Corrosion Risk

    3M is recalling Red Dot Monitoring Electrodes because corrosion could cause performance failures in the ECG monitoring product.

    Product
    3M Red Dot Monitoring Electrode with Foam Tape - Product Usage: intended to be used by healthcare professionals for ECG monitoring. 50 electrodes per pouch; 20 pouches per shipper box (1000 electrodes total).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0377-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 55 mm Number: 00-8065-652-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0381-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 57 mm Number: 00-8065-654-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0360-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Due to Sterilant Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0325-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, Neutral, O.D. Cup with Spacers, 45mm Item Number: 00-8065-542-26 Hip prosthesis component
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0384-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 59 mm Number: 00-8065-656-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0389-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Cups Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the hydrogen peroxide sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 61 mm Number: 00-8065-658-28
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0366-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Component Due to Sterilant Concerns

    Zimmer Biomet is recalling specific ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0340-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 53 mm Item Number: 00-8065-550-32
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0423-2021·2020-11-11

    3M Red Dot Monitoring Electrode Recall Due to Corrosion Risk

    3M is recalling Red Dot Monitoring Electrodes because corrosion could cause performance failures in the ECG monitoring product.

    Product
    3M Red Dot Monitoring Electrode with Foam Tape - Product Usage: intended to be used by healthcare professionals for ECG monitoring. 50 electrodes per pouch; 20 pouches per shipper box (1000 electrodes total).
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0361-2021·2020-11-11

    Zimmer Biomet Recalls Acetabular Cup Hip Prosthesis Components

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. Cup with Spacers, 47 mm Number: 00-8065-644-26
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0357-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling specific acetabular cup hip prosthesis components because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, 10 degree inclined face, O.D. Cup with Spacers, 43 mm Number: 00-8065-640-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0349-2021·2020-11-11

    Zimmer Biomet Recalls Hip Prosthesis Components Due to Sterilant Biocompatibility Concerns

    Zimmer Biomet is recalling 412 units of ZCA All Poly Acetabular Cups because the sterilant used was not evaluated for biocompatibility, posing a potential risk of adverse tissue reactions.

    Product
    ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 22 mm, Neutral, O.D. Cup with Spacers, 59 mm Number: 00-8065-556-22
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-0314-2021·2020-11-11

    GE Healthcare Recalls 9100c NXT Systems Due to Flow Sensor Defects

    GE Healthcare is recalling 37 9100c NXT systems because flow sensors may have damaged tubes with small punctures or cuts.

    Product
    9100c NXT system - Product Usage: Device is licensed and distributed outside of the US only.
    Category
    Medical Device
    Distribution
    36 states